36 - Microsense II Model 5300 Gaging System
User Adjustments
Front Panel Adjustments
Calibration (this is factory-set and should not usually be adjusted)
A scale factor adjustment is available on the front panel. This pot, marked "SF", allows
for fine adjustments to the probe scale factor for periodic calibration. The scale factor is
the displacement-to-voltage ratio.
Zero
A front panel pot is provided to allow fine adjustments of the output voltage offset to
obtain a zero reading without having to move the probe or target. The allowable range is
+10% of the full scale operating range. Adjustment to the Zero pot has no effect on the
Scale Factor.
Limits
Two limit pots are available for setting trip points of the + and - limit lines and LEDs.
Both limit pots may be set anywhere in the range of the probe. The limits are set by
adjusting the probe-to-target distance until the trip point voltage is obtained. The limit pot
is then turned so that the trip point equals the probe voltage. This is defined as the point
at which the LED changes state (ON to OFF or vice versa).