02/25/00
8 - Microsense II Model 5300 Gaging System
Model 5300 Gaging System Specifications
Performance:
Ranges:
±25 microns to ±500 microns full scale, depending on probe selection.
Frequency Response: Jumper selectable filtering: 1 kHz, 5 kHz, 20 kHz, 100 kHz
Linearity:
Typically 0.1% of full-scale (see note 1)
Resolution:
1 nanometer rms @ 100 kHz bandwidth (see note 1)
Functions: (Standard Unit)
Meter: Zero meter reading indicates probe is at nominal standoff from measured target.
Option for differential measurements
A+B:
Sum of the A probe and the B probe displacement signals
A-B:
Difference of the A probe and the B probe displacement signals
EMC Certification:
The Model 5300 Capacitance Gage has been tested and is in compliance with the requirements set forth
in EMC Directives EN 55022, EN 50082-1, 89/336/EC, IEC 801-2, IEC 801-3, and IEC 801-4.
Test Report Number: CE101195
Model 5120 Velocity/Acceleration Option:
A plug-in VA option is available for the Model 5300 system. When configured with this VA option, Velocity
and Acceleration measurements are derived for the probe selected on the channel switch. Filtering is
provided by three-pole Bessel or Butterworth low-pass plug-in filters. The user selects the upper roll-off
frequency (as defined by ANSI standards) over a range from 1-20 kHz.
Me
chanical Parameters:
Dimensions:
Width: 16.5 inches (42.31 cm)
Height: 5.75 inches (14.74 cm)
Depth: 13.0 inches (33.33 cm)
Shipping Weight:
23 pounds (10.35 kg)
Power Requirements:
The 5300 System utilizes a power supply that will accept virtually any standard input voltage in the world.
Please specify the proper IEC320 power cord for your region.
Input Voltages:
100-240 VAC, externally switched
Input Frequency:
50-60 Hz
Notes:
1. Parameters listed are typical when using a 2mm active sensor area and a range of ±50 microns.
Specifications will vary depending upon range, standoff distance, and probe active sensor area. Resolution
improves with larger sensor areas, reduced bandwidth and smaller ranges. Ideal Operating Temperature is
60-90°F (16-35°C). For specific resolution for your application, talk to your ADE Technologies’ sales repre-
sentative.