Microsense II Model 5300 Gaging System - 19
Probe Mounting
Probes are mounted in one of two ways. Axial units are secured by clamping pressure evenly distributed
around the probe tip. Others are mounted with screws inserted into the probe (blade type). Probes have
a preferred clamping area, indicated on the probe outline drawing. Do not use screws to mount probes not
designed to be screw-mounted.
All system fixturing should be rigid and free from any stress that could cause flexing of the surfaces.
Stainless steel or cast-iron fixtures and granite surfaces work well.
The probe should be mounted in a stand that will prevent a collision with the sample. The fixturing stand
should be close enough to the console to allow cable connection. If the probe cable becomes damaged,
it may be easily replaced.
Cleaning Probes
In order to provide accurate measurements, the system probe(s) should be protected from excessive dirt
and dust. Remove foreign materials that may accumulate during normal operation by cleaning the probe's
sensing areas periodically:
n
Use a cotton swab with denatured or isopropyl alcohol to clean the sensing area.
Let dry completely.
When not in use, it is recommended to keep the sensing areas covered with the provided caps.