© Adam Equipment Company 2020
11.2.1.1 Standard output format
The balance will print the following data as the standard form. The standard form cannot
be changed. The format of the custom forms #1 and #2 will be the same as the standard
form until modified by the user.
Line 1
Date
Line 2
Time
Line 3
Blank line
Line 4
ID number
Line 5
Blank line
Line 6
Result
Line 7
Blank line
Line 8
Blank line
This will result in a printout that looks like:
NOTE
: The format of the result line will change depending on the mode in which the
balance is operating, e.g.
Normal weighing, Animal weighing: “
123.456 g
”
Parts counting: “
1234 pcs
”
Percent weighing: “
12.345 %
”
Density: “
12.345 g/cc
”
11.2.2
Custom output format
If output on demand is selected, the user may optionally configure the serial output as a
choice of 3 styles of form, either in a default format or in one of two custom formats. Each
of the custom formats can be configured to output up to 15 lines of data. The data types that
can be printed are:
NAME
TEXT PRINTED
ID number
ID no.: xxxxxxxxxxxx
Serial number
Serial no. xxxxxxxxxxxx
Date
DATE dd/mm/yyyy
Time
TIME hh:mm:ss
Net weight
Net: xxx.xxx g
Gross weight
Gross: xxx.xxx g
Tare weight
Tare: xxx.xxx g
Unit weight
Unit wt: xxx.xxx g
Count
Count: xxxx pcs
Reference weight
Ref. wt: xxx.xxx g
Percent
Percent: xx.xxx %
Date: 01/01/19
Time: 15:45:27
ID No: 123456
Net: 123.456
Содержание Luna LAB 124 e
Страница 2: ... Adam Equipment Company 2020 ...
Страница 16: ... Adam Equipment Company 2020 8 0 KEYPAD The keypad has the following keys to operate the balance ...
Страница 41: ... Adam Equipment Company 2020 18 0 BALANCE MENU STRUCTURE LUNA OPERATING SOFTWARE MENU STRUCTURE ...
Страница 43: ... Adam Equipment Company 2020 Supervisor Level Access continued ...
Страница 46: ... Adam Equipment Company 2020 ...