9.0 TECHNICAL PARAMETERS
The scale has 4 technical parameters.
To set these parameters turn the scale off and then turn it on again. Press the
[Mode]
key during the initial counting from 9 to 0 on the display.
Display will show “
”- the first parameter which can be set by the user.
Press
[Mode]
to scroll through other parameters.
Press
[Zero]
when display shows “
”, next “
"
” will be displayed.
Press
[Smpl]
,
[Print]
and the
[Tare]
key in this order.
Pressing
[Zero]
will allow the user to enter the Technical parameters setting mode.
The first technical parameter “
#
” will be displayed.
Press
[Mode]
to scroll through the technical parameters as below.
#
Corrects Linearity
Allows to enter the Factory Calibration mode. See section
10.0 for details.
Displays the A/D Count to assess range and stability.
The auto-zero will automatically set the scale to zero
when the displayed weight is near zero. This is to make
sure that the scale is measuring weight from an accurate
zero starting point.
The Auto zero value can be set to 0.5, 1, 2 or 4 scale
divisions. The normal setting is 1 scale division. The
display will show “
”.
Содержание ACBplus Series
Страница 24: ...Adam Equipment Company 2005 22 13 0 ACBplus plus plus plus EXPLODED VIEW...
Страница 25: ...Adam Equipment Company 2005 23 14 0 ACBplus plus plus plus MAIN PCB ASSEMBLY...
Страница 26: ...Adam Equipment Company 2005 24 15 0 ACBplus plus plus plus DISPLAY PCB ASSEMBLY...
Страница 27: ...Adam Equipment Company 2005 25 16 0 ACBplus plus plus plus I O PCB ASSEMBLY...
Страница 28: ...Adam Equipment Company 2005 26...