@Adam Equipment Company 2005
13
7.0 ERROR CODES
ERROR
CODES
DESCRIPTION
SUGGESTIONS
Over-range
Remove weight from the scale.
If the problem persists contact your dealer or Adam
Equipment for assistance.
Zero Setting
Error
The scale was outside the normal zero setting range
either when it was turned on or when the
[Zero]
key was
pressed.
Remove weight from the scale and try again.
Use the
[Tare]
key to set the display to zero value.
If the problem persists contact your dealer or Adam
Equipment for assistance.
!
Keyboard Error
Improper operation of the scale.
Incorrect A/D
count
The values from the A/D converter are outside the normal
range.
Remove the weight from the scale if over-loaded. Make
sure the pan is attached.
Indicates the load cell or the electronics may be faulty.
If the problem persists contact your dealer or Adam
Equipment for assistance.
A/D count is out
of normal range
during turning the
scale on
Initial instability during turning on the scale which will get
adjusted automatically.
There may be movement, vibration or dirt on the scale
during turning on.