@Adam Equipment Company 2005
12
6.0 USER CALIBRATION
•
This parameter allows the user to calibrate the scale any time during
the operation.
•
Switch the scale off and then on it again by pressing
[On/Off]
.
•
During the self-test press
[Smpl]
and
[Print]
together.
•
The display will show "
". Remove all weights from the pan.
•
Wait for the stability indicator to be on and then press
[Zero].
•
The display will then show the first calibration weight request.
•
Press
[Mode]
to scroll through the other calibration weights.
•
Press
[Zero]
to select a calibration weight. The display will show
“
”.
•
Place this weight on the platform and press
[Zero]
when the stability
indicator will be on to show the value is stable.
•
Once calibration is complete, the scale will return to normal weighing
counting back to zero. Remove the calibration weight at this stage.
•
Place the weight back on the pan to re-check whether the calibration is
correct. If necessary, repeat the process.
Calibration weights
Model #
ACB
plus
plus
plus
plus
150
ACB
plus
plus
plus
plus
300
ACB
plus
plus
plus
plus
600
ACB
plus
plus
plus
plus
600H
ACB
plus
plus
plus
plus
1000
ACB
plus
plus
plus
plus
1500
ACB
plus
plus
plus
plus
3000
Weight 1
50g
100g
200g
200g
500g
500g
1000g
Weight 2
100g
200g
400g
400g
1000g
1000g
2000g
Weight 3
150g
300g
600g
600g
-
1500g
3000g
Содержание ACBplus Series
Страница 24: ...Adam Equipment Company 2005 22 13 0 ACBplus plus plus plus EXPLODED VIEW...
Страница 25: ...Adam Equipment Company 2005 23 14 0 ACBplus plus plus plus MAIN PCB ASSEMBLY...
Страница 26: ...Adam Equipment Company 2005 24 15 0 ACBplus plus plus plus DISPLAY PCB ASSEMBLY...
Страница 27: ...Adam Equipment Company 2005 25 16 0 ACBplus plus plus plus I O PCB ASSEMBLY...
Страница 28: ...Adam Equipment Company 2005 26...