6.0  USER CALIBRATION  

 

  This  parameter  allows  the  user  to  calibrate  the  scale  any  time  during 

the operation. 

 

  Switch the scale off and then on it again by pressing 

[On/Off]

 

  During the self-test press 

[Smpl] 

and 

[Print]

 together. 

 

  The display will show "

".  Remove all weights from the pan.

 

 

  Wait for the stability indicator to be on and then press 

[Zero]. 

 

  The display will then show the first calibration weight request.   

 

  Press 

[Mode]

 to scroll through the other calibration weights. 

 

  Press 

[Zero]

  to  select  a  calibration  weight.  The  display  will  show 

”. 

 

  Place  this  weight  on  the  platform  and  press 

[Zero]

  when  the  stability 

indicator will be on to show the value is stable. 

 

  Once  calibration  is  complete,  the  scale  will  return  to  normal  weighing 

counting back to zero. Remove the calibration weight at this stage. 

 

  Place the weight back on the pan to re-check whether the calibration is 

correct. If necessary, repeat the process. 

        

 Calibration weights  

 

Model # 

ACB

plus

plus

plus

plus

  

150 

ACB

plus

plus

plus

plus

  

300 

ACB

plus

plus

plus

plus

   

600 

ACB

plus

plus

plus

plus

  

600H 

ACB

plus

plus

plus

plus

  

1000 

ACB

plus

plus

plus

plus

  

1500 

ACB

plus

plus

plus

plus

  

3000 

Weight 1 

50g 

100g 

200g 

200g 

500g 

500g 

1000g 

Weight 2 

100g 

200g 

400g 

400g 

1000g 

1000g 

2000g 

Weight 3 

150g 

300g 

 600g 

600g 

1500g 

3000g 

 

 

Содержание ACBplus Series

Страница 1: ...nt Company 2005 Adam Equipment ACBplus plus plus plus SERIES SERVICE MANUAL Software rev 1 13 for standard models 1 9 for ACBplus 1000 and 2 1H for ACBplus 600H ADAM EQUIPMENT CO LTD p n 6018 Rev B2 F...

Страница 2: ...e revision 1 06 1 07 for standard models and 1 2H 1 3H for ACBplus 600H Manual Revision B1 September 2004 Software rev 1 13 for standard models 1 9 for ACBplus 1000 and 5000 and 2 1H for ACBplus 600H...

Страница 3: ...S 10 5 3 1 Selecting the communication port 10 5 3 2 Setting the print mode 10 5 3 3 Setting the baud rate 10 5 3 4 Parity settings 11 5 4 SETTING OF AUTO SWITCH OFF 11 5 5 TECHNICAL PARAMETERS 11 6 0...

Страница 4: ...then contact your dealer or Adam Equipment for more assistance In order to supply further assistance we will need the following information Name of your company Contact Name Contact telephone e mail...

Страница 5: ...01g 0 01 g 0 05g 0 1g Linearity 0 01g 0 02g 0 04g 0 02g 0 02 g 0 1g 0 2g Pan 120mm 4 7 124 144mm 4 9 X5 7 Wind shield Included as standard Units of Measure g ct Lb OZ d GN OZt dWt MM TL T TL C TL t t...

Страница 6: ...s unstable or Drafts or air currents is displayed Obstruction under pan Sample is moving animal weighing Vibrations through table Temperature changed dramatically Electronics faulty Weight value incor...

Страница 7: ...e for its weight will be displayed Press the Tare key to tare the scale The weight that was displayed is stored as the tare value It is subtracted from the displayed value leaving zero on the display...

Страница 8: ...sample For example is 350g is placed on the scale and the key is pressed the display will show 100 00 Removing the 350g weight and place a 300g weight on the scale The display will show 85 71 as 300g...

Страница 9: ...is set to manual accumulation See the Parameter Section the weight displayed will be stored in memory when the Print key is pressed and the weight is stable The display will show followed by the tota...

Страница 10: ...change the settings of a particular parameter For example to enable the weighing unit pounds change the setting to by pressing Mode now Use the Print key to leave a parameter unchanged The following...

Страница 11: ...ile weighing a sample the display will show the weight and the units of weight currently in use To view the weight in other units which are enabled press Mode 5 2 SETTING OF THE BACKLIGHT The backligh...

Страница 12: ...he print mode is displayed Press Mode to cycle through the other options The following options are available The scale sends the accumulated weights to the communication port whenever the Print key is...

Страница 13: ...ro key to view the current setting for the backlight Press Mode to scroll through other settings 0 5 10 20 and 30 minutes Press Zero to store a particular setting o Use the Print key to leave a parame...

Страница 14: ...display will show Place this weight on the platform and press Zero when the stability indicator will be on to show the value is stable Once calibration is complete the scale will return to normal weig...

Страница 15: ...e problem persists contact your dealer or Adam Equipment for assistance Keyboard Error Improper operation of the scale Incorrect A D count The values from the A D converter are outside the normal rang...

Страница 16: ...Main PCB assembly Battery Display PCB assembly Keypad I O PCB RS 232 Power IN OUT All models of ACBplus are similar except the selection of load cells and the program Normally if a problem is found w...

Страница 17: ...Pressing Zero will allow the user to enter the Technical parameters setting mode The first technical parameter will be displayed Press Mode to scroll through the technical parameters as below Correct...

Страница 18: ...UNT Press Zero when is displayed The A D Count will be displayed If the Load cell is good then Counts should operate between 10 000 and 900 000 with maximum load Press Zero to escape this parameter Pr...

Страница 19: ...key when stable Note The scale should be stable before pressing Zero to accept a weight The stability indicator will be on to show the value is stable Once calibration is complete it will return to no...

Страница 20: ...s the same as pressing Tare key Z cr If Sets the zero point for all subsequent weighing Display shows zero T5 345 cr if Would be same as entering a preset tare value of 5 345 from keypad P cr lf Print...

Страница 21: ...in a particular way on the metal structure These 4 strain gauges are connected in such a way that they form a wheatstone bridge See figure below In the simplest load cells there are only the 4 resista...

Страница 22: ...LY Vout O VOLTS BETWEEN S AND s A method to check the basic function and integrity of a load cell can be done using ohmmeter and voltmeter with up to a 10volt power supply Before the load cell is conn...

Страница 23: ...h 20Kg placed on it Vext is 5 volts would give a output of approximately Vout 2mv V Vext Mass 2mv V 5V 20Kg 6 7mv capacity of load cell 30kg If the load cell has been damaged the no load voltage will...

Страница 24: ...Adam Equipment Company 2005 22 13 0 ACBplus plus plus plus EXPLODED VIEW...

Страница 25: ...Adam Equipment Company 2005 23 14 0 ACBplus plus plus plus MAIN PCB ASSEMBLY...

Страница 26: ...Adam Equipment Company 2005 24 15 0 ACBplus plus plus plus DISPLAY PCB ASSEMBLY...

Страница 27: ...Adam Equipment Company 2005 25 16 0 ACBplus plus plus plus I O PCB ASSEMBLY...

Страница 28: ...Adam Equipment Company 2005 26...

Страница 29: ...to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment The equipment generates uses and can radiate radio frequency energy and if not...

Страница 30: ...tronic Scales for Medical use Retail Scales for price computing Adam Equipment Co Ltd Bond Avenue Milton Keynes MK1 1SW UK Phone 44 0 1908 274545 Fax 44 0 1908 641339 e mail sales adamequipment co uk...

Отзывы: