Parameter
Description
Options
Default
setting
F1 Lo
Set Low limit.
Use the [Unit/
] key
and [Pcs/ ] key to set
the values of the lower
limit.
When set press the
[Z/T] key to store the
value and go to F2 Hi
000000
F2 Hi
Set High limit.
Use the [Unit/
] key
and [Pcs/ ] key to set
the values of the high
limit.
When set press the
[Z/T] key to store the
value and go to F3
bEP
0000000
F3 bEP
This parameter sets the
Beeper to off or on. If set
to on, the beeper can
further be set to sound
when the weighing result
is within or outside the
check-weighing limits.
bP oFF - Off
bP inL - Within limits
bP otL - Outside limits
(>20d)
bP inL
F4 nCK
This parameter enables
negative check weighing
function with ability to do
negative tare.
on
oFF
on
7.2
PERCENT
WEIGHING
AND
ANIMAL
WEIGHING
See section 6.7 and 6.8 for details of these special weighing modes.
•
Press [Z/T] to enter a parameter. Use the [Unit/
] key and [Pcs/ ]
key to set the values or just the [Pcs/ ] key to select the options.
•
Press [Z/T] to confirm the change and then advance to the next
parameter by pressing the [Pcs/ ] key.
•