21
|
P a g e
© Adam Equipment Company 2011
6.8
ANIMAL
(D
YNAMIC
)
WEIGHING
The scale can be set to animal (dynamic) weighing for weighing any items
that are unstable or may move. See Section 7.4.
The scale will use a special filter to minimise the effects of any movement
while the moving animal or unstable samples are on the scale.
•
Press [Func/C]. The first option is “
FunC 1
”, press the [Func/C] key
again to display “
Func 2
”.
•
Press the [Z/T] key. “
F2 PCt
” will be displayed. Press the [Pcs/ ] key to
advance to the second function, “
P2 AnL
”, Animal weighing.
•
Press [Z/T] to enter animal weighing. The scale will change to animal
weighing and allow weighing of an unstable animal or sample on the
platform.
•
To use the Animal Weighing function it is necessary to set the amount
of filtering required for the item to be weighed. More active animals will
require a higher level of filtering to give a stable result. The display will
show “
Flt x
” where x is a value from 1 to 5. The higher the value the
greater the amount of filter will be. To increment the value shown press
the [Pcs/ ] key then press the [Z/T] key to accept it.
•
The display will flash “
Ani
“ 2 times then show the current weight, 0.00.
The scale is now ready to weigh.
Содержание ABK 120