38
|
P a g e
© Adam Equipment Company 2011
12.3
PARAMETER
DETAILS
12.3.1 F1 -CALIBRATION
To enter the calibration parameter, press the [Z/T] key when “
F1
” is displayed.
The scale will be calibrated using 2 masses of approximately 1/3Maximum
and Maximum. If the scale has been calibrated once the values will be
stored. If this is the first time the scale is calibrated the user must enter the
values for the calibration masses.
NOTE: It is necessary to set the decimal point location and the capacity
before calibration is possible.
The display will instruct you to remove any weight from the scale, “
UnloAd
”.
Press [Z/T].
The display will tell you to add the first weight to the scale: “
Ld 1
” “
10 kg
” If
necessary change the value shown to match the weight to be used. Press
[CE] to clear the old value and then enter the new value. All values entered
are in whole numbers only.
Add the weight shown, wait for stability then press the [Z/T] key.
The display will tell you to add the second weight to the scale: “
Ld 2
“ “
30
kg
”
Either add the weight shown, wait for stability then press the [Z/T] key or
press the [Print/M+/Esc] key to accept the single point calibration and return
to weighing.
The display will show “
PASS
” if the calibration is OK.
Remove the weight.
12.3.2 F2–DECIMAL POINT POSITION
To set the value for the decimal point location. The options are 0, 0.0,
0.00, 0.000, 0.0000
To enter this parameter, press the [Z/T] key when “
F2 dEC
” is shown.
The display will show the current setting.
Press the [Pcs/ ] key to change the value. Select from 0, 0.0, 0.00, 0.000,
0.0000
Press [Z/T] to accept the displayed value.