8
|
P a g e
© Adam Equipment Company 2011
3.3
SETTING
UP
THE
SCALES
•
The pillar is attached to the base using a bracket that must be
attached to the base frame first using the 4 bolts supplied. The
pillar is secured to the bracket using 2 sets of screws. The cable
from the base to the indicator module is run through the tube and
taken out through the support at the top. Excess cable can be
stored within the tube.
•
The ABK/AFK Series comes with a stainless steel platform
packed separately. Place the platform in the base.
•
Level the scale by adjusting the four feet. If the scale rocks re-
adjust the feet.
•
Attach the indicator module to the pillar by securing it with the 2
thumb screws provided. Attach the cable from the base to the
connector on the rear of the indicator.
•
Attach the power to the indicator. Press the [On/Off] key. The
software revision number will be displayed followed by a self-test
showing all digits before the zero is displayed along with the unit
of weight that was selected last.