41
|
P a g e
© Adam Equipment Company 2011
12.3.9 F9 –LOW VOLTAGE DETECTION
This parameter allows detection of low voltage when the battery wears down.
To enter this parameter, press the [Z/T] key when “
F9 LVd
” is shown.
The display will show if the LVD Mode is set to on or oFF.
Press the [Pcs/ ] key to change the value.
Press [Z/T] to accept the displayed value.
Press [Print/M+/Esc] to return to weighing.
12.3.10 FA –AZn Auto Zero Range
This parameter sets the auto zero range from 0 (Off) to 4d.
Calibration and parameter counting function to be active. To enter this
parameter, press the [Z/T] key when “
FA AZn
” is shown.
The display will show if the current Auto Zero range.
Press the [Pcs/ ] key to change the value.
Press [Z/T] to accept the displayed value.
Press [Print/M+/Esc] to return to weighing.
12.3.11
Fb –FPS User Function Password
This parameter set a password to control access to the user functions as
described in Section 7. If the password is set to 0000 the control is disable
and the user has free access to the functions.
To enter this parameter, press the [Z/T] key when “
Fb FPS
” is shown.
The display will show “
P1 - - - -
“ Enter the new password code followed by [Z/T]
to accept.
The display will then show “
P2 - - - -
“ Enter the new password code again
followed by [Z/T] to accept. The same code number must be entered both
times for the password to be set.
Press the [Unit/
] and [Pcs/ ] keys to enter the values.
Press [Print/M+/Esc] to return to weighing.