42
|
P a g e
© Adam Equipment Company 2011
13.0
ERROR
CODES
ERROR
CODES
DESCRIPTION
SUGGESTIONS
- -oL - -
Over-range
Remove weight from the scale.
If the problem persists contact your dealer or Adam
Equipment for assistance.
Err 1
Time
Setting
Error
Enter time using correct format and reasonable
values. Format: hh:mm:ss
Err 2
Date
Setting
Error
Enter date using correct format and reasonable
values. Format: yy:mm:dd
Err 4
Zero
Setting
Error
The scale was outside the normal zero setting
range either when it was turned on or when the
[Zero] key was pressed.
Remove weight from the scale and try re-zeroing
again.
Use the [Z/T] key to set the display to zero value.
If the problem persists contact your dealer or Adam
Equipment for assistance.
Err 6
A/D out of range The values from the A/D converter are outside the
normal range.
Remove the weight from the scale if overloaded.
Make sure the pan is attached.
Indicates the load cell or the electronics may be
faulty.
If the problem persists contact your dealer or Adam
Equipment for assistance.
Err 9
Check weigh
limits error
Shown if the low limit is set higher than the current
high limit. Reset High limit or different low limit.
FAIL
Calibration
error.
Improper calibration (should be 10% of the
factory calibration). The old calibration data will be
retained until the calibration process is complete.
If the problem persists contact your dealer or Adam
Equipment for assistance.