6.6
ACCUMULATED
TOTAL
•
The scale can be set to accumulate manually by pressing the
[Print/M+/Esc] key or automatically when a weight is removed
from the scale. See the Section 7.3 for details of setting the
Parameter “
C3 PRM
” and “
C4 Aon
”. The accumulation function is
available when weighing or when counting parts. However the
memory is cleared if the weighing units or functions are changed.
•
When the weight (or count) displayed is stored in memory the
display will show “
ACC 1
” and then the total in memory for 2
seconds before returning to weighing. The RS-232 interface will
output to a printer or PC.
•
Remove the weight, allowing the scale to return to zero and put a
second weight on. When this value is stored, the display will
show “
ACC 2
”, then the new total and finally the value of the
second weight. Repeat as necessary to add all the values needed
to the memory.
•
To view the total in memory press the [Print/M+/Esc] key when
there is no weight on the scale. The display will show the number
of entries and the total.
•
To clear the memory (set the value to zero) press the [Func/C]
key during the time the totals are being displayed.