22
Menu/Enter
In non-programming mode, press this button to enter the
programming mode;
6.4 Button operation descriptions
6.4.1 Button operation of AIM-M300 series insulation monitoring device in RUN mode
(1) Enter RUN the operation mode.
The mode of the default entry is RUN mode, after the LCD displays the software version number, without
any operation, the system goes into RUN mode and starts operation. The main interface shows the
insulation resistance value, temperature value, current value, load rate and system time.
(2) View the alarm records.
In the main interface, press "Down button" to enter the "Fault records query" interface, and press
"enter" button to confirm, turn the pages by pressing "Down button" or "Up button" to query each
fault record in sequence. The first record is the most recent record, and the tenth is the oldest
record.
(3) View software version information.
In the main interface, view the version information of the software by pressing the "Down button"
twice consecutively.
(4) Register address (CAN communication address) to AID200. When AIM-M300 series IMD and AID200
are used together, if AIM-M300(or AIM-M300/SG) doesn’t successfully register address to AID200,
the manual registration is required. In the main interface, press the "Down button" three times
consecutively, entering into the AID200 address registration interface, and press Enter button to
achieve address registration. After the registration, it will automatically return to the main
interface. If the registration is successful, the CAN communication indicator light starts flashing,
indicating that communication is normal.
(5) Device self-test.
In the main interface, press the "TEST" button, then the monitor will start the Self-test program,
simulating the overload fault, insulation fault and over-temperature fault to test whether the
detection function of the device is normal.
6.4.2 Button operation of AIM-M300 series medical insulation monitor in programming mode
(1) Enter the programming mode
In the main interface, press the "Menu" button to enter the password input page of the programming
mode. Input the correct password by pressing "Up " or "Down " button, and press "Enter" button to
enter the programming mode.
(2) Exit programming mode
In the programming mode, press the "Return" key to enter the save confirmation menu, select [Y]
or [N] by the "UP" or "DOWN" key, and then press the "ENTER" key to exit the programming mode. If
[Y] is selected before exit, the parameter settings will be saved; if [N] is selected, the parameters
will not be saved.
(3) System password setting
In programming mode, select [Other Settings] by "Up " or "Down " button, and press "Enter" button