4.2 Technical parameters of AIM-M300 series medical intelligent insulation monitoring device
Refer to Table 3.
Table 3 Technical Parameters of AIM-M300 series Medical Intelligent Insulation Monitoring
AUX Power
Voltage
DC 18…36V
Temperature
monitoring
Thermal
resistor
2 Pt100
Power
consumption
≤6W
Measuring
range
-50…+200℃
Insulation
monitoring
Resistance
measuring
range
15…999kΩ
Alarm value
range
0…+200℃
Response value
50…999kΩ
Alarm
output
Output mode
2 Relays
Relative
uncertainty
±10%,±10K
Contact
rating
AC 250V/3A
DC 30V/3A
Response time
≤3s
Environment
Operating
temperature
-10…+55℃
Permissible
system leakage
capacitance C
e
≤5uF
Transport
temperature
-25…+70℃
Measuring
voltage U
m
≤12V
Storage
temperature
-25…+70℃
Measuring
current I
m
≤5uA
Relative
humidity
5%-95%,No
condensation
Impedance Z
i
≥200kΩ
Altitude
≤2500m
Internal DC
resistance R
i
≥240kΩ
IP degree
IP40
Permissible
extraneous DC
voltage U
fg
≤DC280V
Rated impulse voltage /
pollution degree
4KV/Ⅲ
Load
Measuring Value
2.1…60A
EMC/EMR
IEC 61326-2-4