5.2 Installation
Medical IT system insulation monitoring and fault locating device six pieces of products are
preferably installed in the distribution cabinet (isolation power cabinet) except for the AID150
centralized alarm and display device. The isolation transformer is installed in the bottom of the
cabinet fixed with matching bolts, and the cooling fan should be installed. The device and the circuit
breaker are installed on the upper panel. If the isolation transformer is installed separately,
it is not suitable to put it too far away from the AIM-M300 series insulation monitor. If the AID150
centralized alarm and display device is used in the operation room, it can be embedded in the wall
of the operating room next to the intelligence panel, so that the medical staff can view conveniently.
If it is used in ICU/CCU and other intensive care units, it should be installed in the nurses station,
so that the duty nurses can view. AID150 external wirings include two 24V power cords and a RS485
communication line of two-core shielded twisted pair, which are all drawn from the isolation power
cabinet. Pay attention to reserve pipelines during construction.
5.2.1 Installation of AIM-M300 series medical intelligent insulation monitoring device
AIM-M300 series insulation monitor adopts the installation method of the DIN-rail, and the
fixation mode is the clip buckle type, as shown in the following figure:
5.2.2 Installation mode of HDR-60-24 power module
The HDR-60-24 power module
adopts the installation method of the guide rail, and the fixation
mode is the clip buckle type, which can also be installed on the same guide rail as AIM-M300 insulation