8
Alarm type
Insulation fault, overload, and over temperature
Communication mode
RS485,MODBUS-RTU
Display mode
LCD
display,128*64
dot array
4.5 Technical parameters of HDR-60-24 power supply device
Refer to Table 6.
Table 6 Technical Parameters of HDR-60-24 Power Supply Device
Type
Input
Output
Installation
HDR-60-24
100-240VAC 1.8A
24VDC 2.5A
35mm
DIN-rail
4.6 Technical parameters of AKH-0.66P26 current transformer
Refer to Table 7.
Table 7 Technical Parameters of AKH-0.66P26 Current Transformer
Input current
0.05A~60A
Frequency
range
0.02-10 kHZ
Output current
0.025~25 mA
Loading
resistance
<200Ω
Temperature
coefficient
100 ppm/℃
Transient
current (1s)
200A
Phase
displacement
10
Installation Fixed with 4×10 screws
Operating
temperature
-35~+70℃
Secondary
wiring
Single core >0.75mm
2,
Maximum
length of 1 meter
Storage
temperature
-40~+75℃
Single core twisted pair,
0.75mm
2
,
Maximum length of
10 meters
Secondary
resistance range
95~120Ω
Isolation
pressure
5000Vac
Accuracy
0.5%
Linearity
0.5%
5 Installation and wiring
5.1 Shape and cut-out
5.1.1 Dimensions of AITR series medical isolation transformer (unit: mm)
Dimensions of AITR series medical isolation transformers are shown as below and in Table 9 (unit:
mm)