DETAILED ASSEMBLY
Fluorescence Filter Sets (Fig. 1 & 2)
On the right side of the fluorescence illuminator, loosen the
screw
①
using the M4 Allen key and remove the cover
②
.
Move the fluorescence filter selection lever
③
to the right to
move the filter slider track out
④
for filter cube installation.
Match the dovetail of the first filter cube
⑤
to the groove of the
filter slider channel on the track
④
and push the filter cube
onto the slider channel. Tighten the lock screw for the filter
cube with the M4 Allen key.
To install a second filter cube, follow the instructions above.
Fig. 2 shows the filter cubes correctly installed.
NOTE:
the filter cubes can be installed in any order
according to user preference. You can install up to four (4)
filter cubes
– two (2) on the right side, and two (2) on the
left side of the fluorescence illuminator.
When you are finished installing two blocks on either the right
or left side, slide the filter track back in, replace the cover, and
tighten the screw.
IMPORTANT:
if the lock screw of a filter cube is not
tightened, the filter track will not slide back into position.
The UV Protection Shield (Fig. 3)
Remove the lock screws
①
located on the bottom of the
fluorescence illuminator (Fig. 3). Match up the holes
②
on the
UV protection shield to the lock screw holes then install and
tighten the screws.
The Fluorescence Illuminator (Fig. 4)
Loosen the lock screw completely using the M4 Allen key.
Match the dovetail on the bottom of the fluorescence
illuminator to the dovetail mount on the microscope frame, set
it on top and tighten the lock screw.
Fig. 1
Fig. 2
Fig. 3
Fig. 4