TROUBLESHOOTING 

Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a 
problem occurs, please review the following list and take remedial action as needed. If you cannot solve the 
problem after checking the entire list, please contact your local dealer for assistance. 

Optics 

PROBLEM 

CAUSE

 

SOLUTION 

Although the mercury lamp 
illumination is on, the field of 
view is invisible or dark. 

The light shutter closes the light path 

Switch the light shutter to “O” open 
position 

The ND filter is engaged in the light path. 

Pull out ND filter to open the position 

The fluorescence mirror block is 
improperly engaged in the light path 

Engage it properly 

The aperture iris diaphragm and field iris 
diaphragm are not open enough 

Open the aperture iris diaphragm fully; 
adjust the field iris diaphragm to 
circumscribes the field of view 

Visibility is poor. Image is 
not sharp. Contrast is poor. 

The objective or filter is dirty  

Clean them thoroughly 

The aperture iris diaphragm and field iris 
diaphragm are adjusted improperly 

Open these iris diaphragms fully  

The fluorescence mirror block is not 
proper for the specimen 

Use proper mirror block 

The edge of the field of view 
is obscured or not evenly 
illuminated 
 

The objective is improperly engaged in 
the light path 

Make sure the nosepiece clicks properly 
into place 

The fluorescence mirror block is 
improperly engaged in the light path 

Engage it properly in the light path 

The field of view doesn’t open fully 

Open it fully 

ND filter is stopped in halfway in the light 
path 

Pull in the filter slider until it clicks into 
place 

The mercury lamp is not centered. 

Center it  

The collector focus position is not correct 

Adjust it to an optimum position  

Shadow exists in the field of 
view 

The burner or collector is dusty or stained 

Clean them thoroughly 

 

Electrical System 

The main switch cannot 
supply power to the system 

The power cord is connected improperly 

Connect it properly 

A fuse is blown 

Replace the fuses 

The main switch can be set 
to ON but the burner doesn’t 
ignite 
 

The lamp housing connecting cord is 
connected improperly 

Connect it properly to the connectors 

The mercury lamp is not mounted 

Attach a mercury lamp 

The auto ignition system is 
malfunctioning 

Set the main switch of the power 
supply unit to OFF then on again.  
(Repeated ON-OFF is possible in this 
case) 

The mercury lamp flickers or 
the brightness is low 

The phenomenon is observed in a short 
period after ignition 

Wait for 10 minutes or more after ignition 

The lamp life has expired

 

Replace the mercury lamp 

Содержание 3019

Страница 1: ...MANUAL 73 Mall Drive Commack NY 11725 631 864 1000 P 631 543 8900 F www accu scope com info accu scope com 3019 FLUORESCENCE MICROSCOPE SYSTEM ...

Страница 2: ...6 ASSEMBLY DIAGRAM PROCEDURE 7 10 ADJUSMENT OPERATION INTERPUPILLARY DISTANCE 11 FOCUSING 11 FOCUSING TENSION 11 FIELD DIAPHRAGM 12 APERTURE DIAPHRAGM 12 OIL OBJECTIVE 13 FLUORESCENCE OPERATION PREPARATION 14 ILLUMINATION 15 FILTER FLASHBOARD 15 CENTERING THE MERCURY LAMP 16 CENTERING THE FIELD IRIS DIAPHRAGM 17 SELECTING THE FLUORESCENCE FILTER CUBE 17 TROUBLESHOOTING 18 MAINTENANCE 19 SERVICE 19...

Страница 3: ...an 15 If the bulb is installed at an angle greater than this there is an increased risk of the mercury bulb bursting CARE AND MAINTENANCE 1 Do not attempt to disassemble any component including eyepieces objectives or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More persistent dirt should be r...

Страница 4: ...es ensure each microscope is of the highest quality prior to shipment UNPACKING AND COMPONENTS Your microscope arrived packed in a molded Styrofoam container Do not discard the container the container should be retained for reshipment of your microscope if needed Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew will form Carefully remove t...

Страница 5: ... 73 Mall Drive Commack NY 11725 631 864 1000 www accu scope com 5 COMPONENTS DIAGRAM Lock Screw Y Axis Adjustment Knob Coarse Focusing Knob Fine Focusing Knob X axis Adjustment Knob Power Switch Iris Diaphragm Adjustment Control Objective Nosepiece ...

Страница 6: ...ORESCENCE COMPONENTS DIAGRAM Lock Screw Condenser Adjustment Knob Mercury Bulb Adjustment Screws Mercury Lamp Housing Shutter Slider Field Diaphragm Adjustment Screw Mercury Lamp Power Supply Microscope Frame Fluorescence Filter Selection Lever Fluorescence Illuminator Eyepiece Head Field Diaphragm Adjustment Lever ...

Страница 7: ...ans at our New York facility prior to shipment Should you need to disassemble assemble your microscope in the future please follow the instructions outlined below When assembling the microscope make sure that all parts are free of dust and dirt and avoid scratching any parts or touching glass surfaces Eyepiece Head Mercury Lamp House Mercury Lamp Socket Assembly Fluorescence Illuminator Mercury La...

Страница 8: ... cubes can be installed in any order according to user preference You can install up to four 4 filter cubes two 2 on the right side and two 2 on the left side of the fluorescence illuminator When you are finished installing two blocks on either the right or left side slide the filter track back in replace the cover and tighten the screw IMPORTANT if the lock screw of a filter cube is not tightened...

Страница 9: ...unt the positive pole the wide head before the negative pole or damage to the lamp may occur Never use excessive force when mounting the mercury lamp Avoid touching or soiling the mercury lamp If this happens gently wipe it with lens tissue or a soft clean cloth To prevent hazard always turn the main switch on the power supply unit to O OFF position unplug the power cord plug from the main outlet ...

Страница 10: ...e mercury lamp house and tighten the lock screw Connect the other side of the plug to the port on the mercury lamp power supply Connect the one side of the power cord to the socket on the mercury power supply and the other to an electrical outlet that is well grounded The Fuse Fig 11 Make sure the main switch to the microscope and mercury power supply are in the off O position Remove the plug Fast...

Страница 11: ...ng manner set both diopter collars to 0 Using your left eye only and the 10X objective focus your specimen by adjusting the coarse adjustment knob When the image is in view refine the image to its sharpest focus by turning the fine adjustment knob Rotate the diopter collar to obtain the sharpest focus To obtain the same sharp image using your right eye do not touch the coarse or fine adjustments I...

Страница 12: ... the best performance and obtain the clearest image Fig 15 Adjusting the Aperture Diaphragm The aperture size is increased or decreased by rotating the condenser aperture diaphragm lever When the aperture is closed the brightness and resolution are decreased but the contrast and range of focus are increased If the aperture diaphragm is opened the brightness and resolution are increased however the...

Страница 13: ...ctive back and forth Rotate the nosepiece so the 100xR oil immersion objective is in the light path With your eye at the level of the stage use the coarse focus knob to raise the stage with the specimen cover glass When you see a flash of light at this location the objective lens has made contact with the immersion oil and the microscope can now be focused using the fine focus knob Each time you f...

Страница 14: ...ff of the mercury lamp will shorten its service life considerably Precautions on the specimen color fading The system employs high intensity excitation light to enable bright observation of dark fluorescence specimens As a result if high power objectives are used frequently color fading of the specimen occurs early degrading the view contrast of fluorescence images So it is effective to use the sh...

Страница 15: ...s from the timer LIFE TIME press and hold the reset button RESET for more than 5 seconds The indicating range of the current CURRENT is 0 9 99A No need to switch on for transmitted illumination when using fluorescence illumination Do not cut off power supply within 15 minutes after mercury bulb is triggered to avoid damage to the bulb In order to prolong the life of mercury bulb please do not re l...

Страница 16: ...g knob of mercury bulb and the horizontal adjusting knob of mercury bulb to make the bulb image on the scale of the slide Fig 22 A Adjust the knob of the parabolic mirror to focus the horizontal centering knob and the vertical centering knob to make the bulb mirror image on the scale of the slide Fig 22 B Adjust the knob and to make the bulb image and bulb mirror image symmetrical to the scale of ...

Страница 17: ...ragm lever out until the diaphragm comes into the smallest state Use the two centering screws to adjust the two field iris diaphragm centering screws alternately to move the image of the diaphragm to the center Fig 24 shows the adjustment of diaphragm Move the field diaphragm lever to open the diaphragm As this makes slight deviation noticeable adjust the centering precisely Enlarge the diaphragm ...

Страница 18: ...s not proper for the specimen Use proper mirror block The edge of the field of view is obscured or not evenly illuminated The objective is improperly engaged in the light path Make sure the nosepiece clicks properly into place The fluorescence mirror block is improperly engaged in the light path Engage it properly in the light path The field of view doesn t open fully Open it fully ND filter is st...

Страница 19: ...a period of five years from the date of invoice to the original end user purchaser The mercury power supply is warranted for a period of one year from the date of invoice to the original end user purchaser This warranty does not cover damage caused in transit misuse neglect abuse or damage resulting from improper servicing or modification by other then ACCU SCOPE approved service personnel This wa...

Отзывы: