50
Situation
Cancelling the Temporary Basal Rate Result
When programming a TBR
j
If you do not press any meter buttons
for approximately 20 seconds, the meter
returns to the Pump Run screen.
j
Return to the TEMPORARY BASAL RATE
(TBR) screen by pressing
.
or by pressing
<
and
x
simultaneously.
j
Set the TBR percentage to 100 %. Press
>
to exit the TBR PERCENTAGE screen.
The pump continues to
deliver the normal basal
rate.
During the delivery of the TBR
j
Press
.
. The meter beeps and returns to
the Pump Run screen.
j
You can put the pump in Stop mode.
The TBR delivery is cancelled as well as
any Extended or Multiwave bolus that is
currently running. The meter displays the
pump warning W6: TBR CANCELLED. Press
>
twice to confirm and dismiss the warning.
The TBR is cancelled and
the pump delivers the
normal basal rate.
For more information, see the pump User Guide.
Cancelling a Temporary Basal Rate