Display Shows
Possible Cause(s)
Possible Solution(s)
Communication Lost
j
Pump is out of range of the meter.
j
Review or cancel the bolus on the pump.
j
Ensure the meter and pump have
Bluetooth
wireless technology turned on.
j
Ensure the meter and pump are within
communication range.
j
If the low battery icon is displayed on the
meter, the meter cannot communicate
with the pump. Replace the batteries.
Communication Warning
j
Communication of bolus data from
the pump was not successful.
Therefore, pump data are not
available and the bolus data may
not be accurate.
j
Communication of bolus data from
the pump was not successful.
Therefore, pump data are not
available and the active insulin
amount may not be accurate.
j
Ensure the meter and pump have
Bluetooth
wireless technology turned on.
j
Ensure the meter and pump are within
communication range.
j
If the low battery icon is displayed on the
meter, the meter cannot communicate
with the pump. Replace the batteries.