77
5
Display Shows
Possible Cause(s)
Possible Solution(s)
Infrequent Pump
Communication
j
At least two weeks have passed
since the meter and pump have
communicated.
j
It is important to use
Bluetooth
wireless
technology communication between the
meter and pump regularly if you utilize
bolus advice.
j
Ensure the meter and pump have
Bluetooth
wireless technology turned on.
j
Ensure the meter and pump are within
communication range.
j
If the low battery icon is displayed on the
meter, the meter cannot communicate
with the pump. Replace the batteries.
Invalid PIN
j
The wrong PIN was entered.
j
Select OK and re-enter the PIN shown on
the pump display.
45676_adv.indb 77
11/3/09 4:45:22 PM
Содержание Aviva
Страница 2: ...LAST UPDATE 2009 07 45676_adv indb 2 11 3 09 4 45 05 PM...
Страница 6: ...4 45676_adv indb 4 11 3 09 4 45 05 PM...
Страница 8: ...6 45676_adv indb 6 11 3 09 4 45 05 PM...
Страница 26: ...24 45676_adv indb 24 11 3 09 4 45 10 PM...
Страница 34: ...32 45676_adv indb 32 11 3 09 4 45 10 PM...
Страница 72: ...70 45676_adv indb 70 11 3 09 4 45 22 PM...
Страница 110: ...108 U upper blood glucose target limit 84 W warning limits 83 warnings pump 53 54 45676_adv indb 108 11 3 09 4 45 26 PM...