76
Display Shows
Possible Cause(s)
Possible Solution(s)
Connection Lost
j
The connection was lost between
the meter and the pump during the
pairing process. Therefore, the
attempt to pair your meter and
pump was unsuccessful.
j
Restart the pairing process. See Chapter
4, “Communication Between the
Accu-Chek Aviva Combo Meter and
Accu-Chek Spirit Combo Pump.”
j
Ensure the meter and pump have
Bluetooth
wireless technology turned on.
j
If the low battery icon is displayed on
the meter, the meter cannot
communicate with the pump. Replace
the batteries.
j
Pump is out of range of the meter.
j
Ensure the meter and pump are within
communication range.
Incorrect Pump Time/Date
j
Pump time/date is incorrect.
j
Set time/date on the pump (see pump
User Guide).