![background image](http://html1.mh-extra.com/html/accu-chek/aviva/aviva_advanced-owners-booklet_3745949076.webp)
74
Display Shows
Possible Cause(s)
Possible Solution(s)
Bolus Delivery Unavailable
j
The meter cannot communicate
with the pump.
j
Retry or set to Manual Pump.
j
Ensure the meter and pump have
Bluetooth
wireless technology turned on.
j
Ensure the meter and pump are within
communication range.
j
If the low battery icon is displayed on
the meter, the meter cannot
communicate with the pump. Replace
the batteries.
j
The pump is currently in Stop
mode.
j
Take the pump out of Stop mode.
j
The pump is currently delivering a
bolus.
j
Allow the current bolus delivery to
complete before delivering the next
bolus.
45676_adv.indb 74