background image

IN4 Care

By IN4 Technology

GlucoAlpha

 

ENHANCE

Blood Glucose Monitoring System

User Guide

P30-A311

IN4 Technology Corporation

www.in4-technology.com

About GlucoAlpha™ ENHANCE 

INTENDED USE

The GlucoAlpha™ Enhance Blood Glucose Monitoring System 
(The GlucoAlpha™ Enhance System) is designed to 
quantitatively measure the concentration of glucose in 
capillary whole blood by persons with diabetes for self-
testing or by healthcare professionals for monitoring glucose 
in the home or in healthcare facilities. The device is 
indicated for professional use and over-the-counter sale. 

ABOUT THE SYSTEM

The GlucoAlpha™ Enhance System uses the latest technology 
in blood glucose monitoring to provide you with easy and 
comfortable testing. The system requires only a 1.5

μ

L of 

blood sample to complete the testing in only 15 seconds. 

The GlucoAlpha™ Enhance System consists of       
1) GlucoAlpha™ Enhance Blood Glucose Meter,   
2) GlucoAlpha™ Blood Glucose Test Strips,  
3) GlucoAlpha™ Check Strip     
4) GlucoAlpha™ Control Solution. 

These products are intended to be used together to get 
accurate blood glucose test results. Do not use other test 
strips or control solutions with your meter.

IMPORTANT INFORMATION

 

        GlucoAlpha™ Enhance System is intended for in vitro 
diagnostic use with capillary whole blood. The system should 
not be used for diagnosis of diabetes or for testing newborn 
infant (neonatal testing). 

     CAUTION

n

The user should not take any decision of medical 
relevance without first consulting his or her medical 
practitioner.

n

Call your doctor immediately if you experience symptoms 
that are not consistent with your blood glucose test results.

n

Severe dehydration or excessive water loss may cause 
false, low results. Call your doctor right away if you 
believe you are suffering from dehydration.

n

A sample with large amount of reducing substances such as 
ascorbic acid and uric acid may cause result slightly higher 
than the actual glucose level.

n

A red blood cell count (hematocrit) that is either very high 
(over 55%) or very low (under 30%) can cause false result.

n

High altitudes above than 2,750 meter (8,800 feet) may 
affect the test results. 

n

Temperatures outside the range of 10°C to 40°C (50°F to 
104°F) may affect the test results.

GlucoAlpha™ ENHANCE BLOOD GLUCOSE METER

Study this diagram and become familiar with all parts of your 
GlucoAlpha™ Enhance Blood Glucose Meter.

1.

DISPLAY

: The large, easy to read display shows blood 

glucose results, messages, glucose results stored in 
memory, time and date.

2.

S BUTTON

: Press S button to enter code setting and when 

you need to decrease values in setting mode.

3.

M BUTTON

: Press M button to enter memory recall mode 

and to increase values in setting mode.

4.

STRIP SLOT

: Holds a GlucoAlpha™ Blood Glucose Test Strip 

and Check Strip in place when you perform blood glucose 
test or perform check test.

5.

BATTERY COMPARTMENT

: Holds ONE 3v Lithium battery 

(CR2032). The battery is not yet installed into meter when 
new purchasing. By using the meter, please install the 
battery first. Refer to Section “TAKING CARE OF YOUR 
METER” of the manual for more information.

GlucoAlpha™ ENHANCE LCD WINDOW

1.

TEST RESULT

2.

BATTERY SYMBOL

: Appears when battery is weak.

3.

THERMOMETER SYMBOL

: Appears when ambient 

temperature is above or below the acceptable range 
needed for testing.

4.

BLOOD SYMBOL

: Appears when the test strip is inserted 

and meter is ready for blood testing.

5.

STRIP SYMBOL

6.

M188

: Memory capacity is 153 entries. The display shows 

M188 to ensure that all digits are working properly.

7.

mg/dL mmol/L

: The meter has two units of measurement 

to choose from.

8.

DAY AVG

: Appears when the meter is in Memory Mode. The 

meter automatically calculates your 7/14/30 days before 
meal (AC) and after meal (PC) test averages if you preset 
correct time and date.

9.

DATE

: Month/date.

10.

TIME

: The meter is programmed for a 12 hours period format.

4

2

1

3

5

5

4

10

6

3

2

1

7

9

8

GlucoAlpha™  
BLOOD GLUCOSE TEST STRIP

1.

Aperture

: apply the blood from either side of strip.

2.

Test Confirmation Window

: make sure the part is filled 

up with blood to ensure the correct result.

3.

Gray Electrode End

: insert this end of test strip into meter.

      IMPORTANT TEST STRIP INFORMATION

1. GlucoAlpha™ Enhance System measures the amount of 

glucose in capillary whole blood. Blood can be applied 
both to the right or left side of the test strip's apertures 
and is automatically drawn to the test confirmation 
window where the reaction takes place.

2. GlucoAlpha™ Blood Glucose Strips are intended for in vitro 

diagnostic use with capillary whole blood or GlucoAlpha™ 
Control Solution. Results will not be accurate if used with 
plasma or serum samples.

3. Do not use glucose strips beyond the expiration date 

indicated on the strip vial label. 

4. The discard date for glucose strips is 90 days after first 

opening the vial. Record the discard date on the vial, 
when you open a new vial of glucose strips.

5. GlucoAlpha™ Blood Glucose Strip can be damaged by heat 

and light. Keep them sealed in the original vial. 

6.                       Store the vial in a cool, dry place below to 

40

˚

C/104

˚

F and above to 10

˚

C/50

˚

F. Do not refrigerate. 

7. Do not use damaged glucose strips or glucose strips 

changed in any way. Use glucose strip immediately after 
removing it from the vial; replace the vial cap and close it 
tightly.

8. Do not transfer glucose strips to a new vial. Always carry 

glucose strips in their original vial.

9. Do not place in direct heat or sunlight.

10. Do not carry loose glucose strips in your carrying case.

11.       Blood glucose test strips are for single use only.

12. Do not use fluoride or iodoacetic acid as a preservative 

for blood specimens.

1

1

2

3

+10

˚

C

(50

˚

F)

+40

˚

C

(104

˚

F)

GlucoAlpha™ CHECK STRIP 

The GlucoAlpha™ Check Strip can be used in 2 ways:

1. To check that the meter is operating 

properly. Use when performing 
quality control check on your meter.

2. To change the measuring unit of 

your meter.

HOW TO CHECK METER BY CHECK STRIP

Step 1

: Insert the check strip into strip slot with label side 

up as above.

Step 2

: You should obtain an acceptable “OK“ reading within 

3 seconds, which mean your meter is working properly. 

Step 3

: Remove the check strip to exit. Meter will 

automatically off.

NOTE

: IF YOU DO NOT GET “OK” READING BUT APPEAR OTHER 

ERROR MEEAGE, REMOVE CHECK STRIP TO TURN OFF THE METER 
FIRST. THEN REPEAT THE TEST. IF THE SECOND RESULT PERSISTS, 
PLEASE CONSULT YOUR LOCAL DISTRIBUTOR FOR HELP.

How To Change Measuring Unit

The meter has two measuring units to choose from, mg/dL 
and mmol/L. The default unit is mg/dL if you do not change. 
Consult your doctor for the proper unit in your country.

Step 1

: Insert check strip into strip slot with label side up.

Step 2

: After “OK” displayed, press S button for 3 seconds 

until flashing mg/dL shows up.

Step 3

: Press S button to switch from mg/dL to mmol/L. 

Wait the new setting flash 4 times to validate the 
new setting. Meter will display “OK” before OFF, and 
the unit is been changed successfully.

Step 4

: Remove the check strip from the meter.

SETTING TIME AND DATE 

Time and date should be set prior to testing if you plan to 
review your 7/14/30 days test average in before meal (AC) 
and after meal (PC) reading or if you want to download the 
results into computer or to review your memory to check 
when a test was done.

TO SET THE TIME AND DATE

1. Make sure the meter is turned off. 
2. Let the front of the meter rest in the palm of your hand. 
3. Slide battery compartment door open and you will find a 

“RESET” hole above the battery.

4. Prink the “RESET” hole to enter time and date setting. 
5. At the beep sound, the year will flash - for example for 

the year 2006, number “06” will appear.

6. Press S or M BUTTON to obtain the desired year.
7. Wait the new setting flash 4 times to confirm and move to 

next setting.

8. Repeat to set the month, day, hour and minute. 
9. After minute is set, the meter displays OK and turn off 

automatically.

CODING YOUR METER 

Before you use your meter for the first time and every time 
you open a new box of test strips, you need to set the meter 
to “match” the strips. This is called coding.

1. Insert the test strip into the strip slot (FIG 1) to turn on 

the meter. Then the last code number will appear on the 
screen.

2. Compare the code number shown on the meter display 

against the code number on the test strip vial (FIG 2). If 
the two numbers match you may begin blood testing. 
Otherwise continue to the next step.

3. Press S button to enter the code mode and the code value 

flashes (FIG 3).

4. Press S or M button to obtain the code number indicated 

on the new test strip vial (FIG 4).

5. Upon obtaining the right code, waiting new setting flash 4 

times to validate the change. After screen shows the 
proper code and a blinking blood symbol shows up (FIG 5), 
your meter is ready to perform blood test.

NOTE: when the meter is OFF, you can also press and S button for 3 seconds to code the meter directly. The coding procedures 
are the same as above step 4 and 5.

SET LANCING DEVICE 

1. Remove the depth adjustment cap. Insert a lancet into the 

lancet holder and push down until it is fully seated (FIG 1).

 

2. Twist off the protective cap until it separates from the 

lancet (FIG 2). 

3. Replace the depth adjustment cap and set the puncture 

depth to the desired number (FIG 3). 

NOTE: The Larger The Number Setting The Deeper The Puncture.

 

4. Pull back the cocking control until it makes a click, and 

then release (FIG 4). If it does not click, the device may 
have been cocked when the lancet was inserted.

PREPARE FOR BLOOD SAMPLING 

Your lancing device and lancets are used for obtaining 
capillary blood samples from the fingertip.

ADJUSTABLE LANCING DEVICE

1. Depth Adjustment Cap

2. Depth Selector 

3. Lancet Holder

4. Cocking Control

5. Release Button

Important

n

The lancing device is in conformity with MDD 93/42/EEC. 
Refer to package insert for contact information of 
manufacturer and CE marking.

LANCET

Protective cap 

Important

n

       Lancets are for single use only. 

n

Keep lancing device and lancets clean.

n

Use caution when removing the used lancet from the 
device and when disposing the used lancet.

n

The lancets are in conformity with MDD 93/42/EEC. Refer 
to product labels for contact information of manufacturer 
and CE marking.

Reviews: