10.   Withdraw the control fluid slowly into a 1 or 3 cc plain 

syringe until the syringe is half filled.  Dispense 3 drops  
of control fluid onto a gauze pad.

Draw fluid

Fill cartridge

Fold closure

Control Testing Procedure - continued  

Review Results

•  Compare results on handheld to the Value Assignment 

Sheet. 

•  If any result is out of range, repeat the control testing with 

a fresh ampule and syringe.

•   If any result is still out of range, do not use the cartridge 

lot, and call Technical Support at 800-284-0702,  
Option 1.

11.   Fill the cartridge with the control fluid. Fold the  

closure over the sample well until it snaps into place.

12.   Push cartridge into the handheld port until it clicks into place, and do not remove until the 

“CARTRIDGE LOCKED” message is gone.   

Содержание i-STAT 1

Страница 1: ...Quick Reference Guide for the i STAT 1 Handheld and i STAT Chemistry Cartridges For use with a CLIA Certificate of Waiver CHEM8 Crea G E3 EC4 6...

Страница 2: ...instructions You should read the complete test procedure before performing the test See the i STAT System Manual for Waived Tests for complete information Individual Cartridge Must be at room tempera...

Страница 3: ...trol Level 1 control EC4 E3 Glucose Cartridges use the i STAT or TriControl Level 3 control Test Control Control Testing Procedure Shake ampule Break ampule MENU 1 3 1 Press to turn on handheld 2 Pres...

Страница 4: ...o the Value Assignment Sheet If any result is out of range repeat the control testing with a fresh ampule and syringe If any result is still out of range do not use the cartridge lot and call Technica...

Страница 5: ...handheld 2 Press i STAT Cartridge 3 As prompted by the handheld use number keys to enter the Operator ID and Patient ID 4 Scan the barcoded cartridge lot number 5 Remove cartridge from pouch 6 Use a...

Страница 6: ...losure Insert cartridge Review Results View results on the handheld Report results as per your facility policy Patient Testing Procedure continued 10 Push the cartridge into the handheld port until it...

Страница 7: ...nd box Refrigerated Storage 2 8 C 35 to 46 F Room Temperature Storage Follow instructions on the cartridge pouch and box Control Storage Do not use controls after the expiration date on the labeling R...

Отзывы: