9. Fold the snap closure over the sample well until it clicks into place.
Press on the tab. Do not press directly over the sample well.
Pt: 12345
16:10 25JanYY
i-STAT CHEM8+
Na
mmol/L
141
K
mmol/L
4 . 2
Cl
mmol/L
1 0 4
iCa
mmol/L
1 . 2 2
T C O 2
m m o l / L
2 6
Page
Pt: 12345
16:10 25JanYY
i-STAT CHEM8+
Glu
mg/dL
90
BUN
mg/dL
1 7
Crea
mg/dL
1 . 0
Hct
%PCV
4 2
H b
g / d L
1 4 . 3
AnGap
mmol/L
16
Page
If all results are not displayed on the first
screen, press to page through all
screens.
Fold closure
Insert cartridge
Review Results
• View results on the handheld.
Report results as per your
facility policy.
Patient Testing Procedure - continued
10. Push the cartridge into the handheld port until it clicks into place.
Do not remove the cartridge from the handheld until the
“CARTRIDGE LOCKED” message is gone. For Hct testing, the
handheld should remain level until a result is obtained. A level surface
includes running the handheld in the downloader/recharger.
• See
i-STAT 1 System
Manual for Waived Tests
if the results are not
displayed or if results are
questioned.
11. After reviewing results, discard gloves, tube, transfer device and cartridge in
biohazard waste container.