5
Prepare the Sample
• Mix the sample tube by gently inverting it 2-3 times before filling a
cartridge. Do not use a partially filled collection tube.
•
CHEM8+
: Test the sample within
10 minutes
after collecting it.
•
6+
,
EC4+
,
E3+
,
Crea
, and
G
: Test the sample within
30 minutes
after collecting it.
Patient Testing Procedure
1.
Press to turn on handheld.
2.
Press (i-STAT Cartridge).
3. As prompted by the handheld, use number
keys to enter the Operator ID and Patient ID.
4. Scan the barcoded cartridge lot number.
5. Remove cartridge from pouch.
6. Use a sample transfer device to remove some
sample from the tube. Dispense 3 drops of
sample onto a gauze pad.
7. Hold the sample transfer device over the
cartridge sample well.
Test Sample
Get sample
Fill cartridge
2
8. Dispense sample until blood reaches the
cartridge fill mark. Some blood must be left in
the cartridge well.
Continued