manualshive.com logo in svg
background image

                                                                  

Quick Reference Guide for the i-STAT

®

 

Handheld and i-STAT Chemistry Cartridges

For use with a CLIA Certificate of Waiver

 

This Quick Reference Guide is for use with the following CLIA-waived i-STAT 
cartridges:  

CHEM8+

6+

EC4+

E3+

Crea

, and 

G

.  These cartridges include 

various subsets of the following tests: sodium, potassium, chloride, total carbon 
dioxide, ionized calcium, glucose, urea nitrogen, creatinine and hematocrit.  

 

CLIA-waived status applies to testing performed on venous whole blood collected in 
tubes with lithium heparin anticoagulant only.

Your facility must have a CLIA Certificate of Waiver in order to perform waived tests, 
must meet all applicable state and local laboratory testing laws, and must follow the 
manufacturer’s instructions.

You should read the complete test procedure before performing the test.  See the 
i-STAT System Manual for Waived Tests for complete information.

• 

 Individual  Cartridge

:  Must be at room temperature for at least 

5 minutes

 before using it.

•  

Box of Cartridges

: Must be at room temperature for at least 

1 hour

 before using the cartridges.

•  Do not remove the cartridge from the pouch until instructed to in the procedures.

Prepare the Cartridge

To scan the barcoded cartridge lot number, position the barcode 3 - 9 inches from laser barcode scanner 
window on the handheld.  Press and hold            to activate the scanner. Align the red laser light so it covers 
the entire barcode. The handheld will beep when it reads the barcode successfully. 

SCAN

Tear open pouch at notch.

Remove cartridge.  
Always hold by sides

.

Place on level surface.

 

Scan the barcode.

Fill Mark

Tab for Snap Closure

Sample Well  

(Blood or control fluid 

goes here)

Cartridge Features

Содержание i-STAT 1

Страница 1: ...Quick Reference Guide for the i STAT 1 Handheld and i STAT Chemistry Cartridges For use with a CLIA Certificate of Waiver CHEM8 Crea G E3 EC4 6...

Страница 2: ...instructions You should read the complete test procedure before performing the test See the i STAT System Manual for Waived Tests for complete information Individual Cartridge Must be at room tempera...

Страница 3: ...trol Level 1 control EC4 E3 Glucose Cartridges use the i STAT or TriControl Level 3 control Test Control Control Testing Procedure Shake ampule Break ampule MENU 1 3 1 Press to turn on handheld 2 Pres...

Страница 4: ...o the Value Assignment Sheet If any result is out of range repeat the control testing with a fresh ampule and syringe If any result is still out of range do not use the cartridge lot and call Technica...

Страница 5: ...handheld 2 Press i STAT Cartridge 3 As prompted by the handheld use number keys to enter the Operator ID and Patient ID 4 Scan the barcoded cartridge lot number 5 Remove cartridge from pouch 6 Use a...

Страница 6: ...losure Insert cartridge Review Results View results on the handheld Report results as per your facility policy Patient Testing Procedure continued 10 Push the cartridge into the handheld port until it...

Страница 7: ...nd box Refrigerated Storage 2 8 C 35 to 46 F Room Temperature Storage Follow instructions on the cartridge pouch and box Control Storage Do not use controls after the expiration date on the labeling R...

Отзывы: