Control Testing Frequency

• Test one cartridge from each lot in each shipment upon receipt.
•  Test a single cartridge from the refrigerator monthly. Select cartridge in the following order: 

CHEM8+, 6+, Crea, EC4+, Glucose, E3+

Prepare Control

• 

 Control  Ampule

:  Must be at room temperature for at least  

 

 

          

30 minutes

 before using it.

•  

CHEM8+ Cartridges

:  use the TriControl Level 1 control.

• 

 6+ & Crea Cartridges

: use the i-STAT or TriControl Level 1  

 

 

 

     control.

• 

 EC4+, E3+, Glucose Cartridges

:  use the i-STAT or TriControl 

Level 3 control.

Test Control

Control Testing Procedure 

Shake ampule

Break ampule

MENU

1

3

1. 

 

 Press            to turn on handheld.

2.  Press            to change screen to Administration Menu.

3.  Press            for the Quality Tests menu.

4.  Press            for Control.

5.   As prompted by the handheld, scan or manually enter the 

Operator ID and Control Lot number.

6.   Scan the barcoded cartridge lot number.  

7.  Remove cartridge from pouch.

8.   Shake the ampule hard for 10 seconds.

9.   Break the ampule open. Use an ampule breaker or 

protect fingers by covering the ampule with gauze.

Continued

Содержание i-STAT 1

Страница 1: ...Quick Reference Guide for the i STAT 1 Handheld and i STAT Chemistry Cartridges For use with a CLIA Certificate of Waiver CHEM8 Crea G E3 EC4 6...

Страница 2: ...instructions You should read the complete test procedure before performing the test See the i STAT System Manual for Waived Tests for complete information Individual Cartridge Must be at room tempera...

Страница 3: ...trol Level 1 control EC4 E3 Glucose Cartridges use the i STAT or TriControl Level 3 control Test Control Control Testing Procedure Shake ampule Break ampule MENU 1 3 1 Press to turn on handheld 2 Pres...

Страница 4: ...o the Value Assignment Sheet If any result is out of range repeat the control testing with a fresh ampule and syringe If any result is still out of range do not use the cartridge lot and call Technica...

Страница 5: ...handheld 2 Press i STAT Cartridge 3 As prompted by the handheld use number keys to enter the Operator ID and Patient ID 4 Scan the barcoded cartridge lot number 5 Remove cartridge from pouch 6 Use a...

Страница 6: ...losure Insert cartridge Review Results View results on the handheld Report results as per your facility policy Patient Testing Procedure continued 10 Push the cartridge into the handheld port until it...

Страница 7: ...nd box Refrigerated Storage 2 8 C 35 to 46 F Room Temperature Storage Follow instructions on the cartridge pouch and box Control Storage Do not use controls after the expiration date on the labeling R...

Отзывы: