Cartridge Receipt

•  Check the temperature strip included with each shipment.   

Do not use the cartridges if the 3rd or 4th windows on the strip are colored.  
Contact Technical Support at 800.284.0702, option 1.

Cartridge Storage

•  Do not use the cartridges after the expiration date on the cartridge pouch and 

box.

• 

 Refrigerated  Storage

:  2-8 ˚C (35 to 46 ˚F)

•  

Room Temperature Storage

:  Follow instructions on the cartridge pouch 

and box.

Control Storage

•  Do not use controls after the expiration date on the labeling. 

• 

 Refrigerated  Storage

:  2-8 ˚C (35 to 46 ˚F)

•  

Room Temperature Storage

:  18-30 ˚C (64 to 86 ˚F) for up to 5 days.

Cartridge and Control Storage

Precautions

To protect yourself and others from infection, use standard precautions 
as defined by your  organization or by the Occupational Safety and 
Health Administration (OSHA).

A falling or dropped handheld: place handheld and peripherals on a stable 
surface.

Barcode scanner: do not look into laser beam coming from scanner, or 
point into eyes of someone else.

Potential Sources of  
Harm to the Operator

Содержание i-STAT 1

Страница 1: ...Quick Reference Guide for the i STAT 1 Handheld and i STAT Chemistry Cartridges For use with a CLIA Certificate of Waiver CHEM8 Crea G E3 EC4 6...

Страница 2: ...instructions You should read the complete test procedure before performing the test See the i STAT System Manual for Waived Tests for complete information Individual Cartridge Must be at room tempera...

Страница 3: ...trol Level 1 control EC4 E3 Glucose Cartridges use the i STAT or TriControl Level 3 control Test Control Control Testing Procedure Shake ampule Break ampule MENU 1 3 1 Press to turn on handheld 2 Pres...

Страница 4: ...o the Value Assignment Sheet If any result is out of range repeat the control testing with a fresh ampule and syringe If any result is still out of range do not use the cartridge lot and call Technica...

Страница 5: ...handheld 2 Press i STAT Cartridge 3 As prompted by the handheld use number keys to enter the Operator ID and Patient ID 4 Scan the barcoded cartridge lot number 5 Remove cartridge from pouch 6 Use a...

Страница 6: ...losure Insert cartridge Review Results View results on the handheld Report results as per your facility policy Patient Testing Procedure continued 10 Push the cartridge into the handheld port until it...

Страница 7: ...nd box Refrigerated Storage 2 8 C 35 to 46 F Room Temperature Storage Follow instructions on the cartridge pouch and box Control Storage Do not use controls after the expiration date on the labeling R...

Отзывы: