Example 2: Maintenance Request menu - overview table
Maintenance request - Error
Probable Cause
Measures
T01 Probe Heat TAH
Probe sample tube heating temperature high
Check process and water temperature
T01 Probe Heat TAL
Probe sample tube heating temperature low
Check process and water temperature
T02 Heated hose TAH
Heated hose temperature high
Check heated hose connection and hose for damages
T02 Heated hose TAL
Heated hose temperature low
Check heated hose connection and hose for damages
T03 Heated valve TAH (Option)
Heated valve temperature high
Check heated valve connection and hose for damages
T03 Heated valve TAL (Option)
Heated valve temperature low
Check heated valve connection and hose for damages
T10 Water tank TAL
Water tank temperature low
Check process and ambient temperature and if anti-freeze has
been added
L11 Water tank LAL
Water tank level low
Check refill valve of potable water to the tank
F12 Water flow FAL
Water flow low
Check water system for leakages and lime deposits
T01 Ai Error
Probe sample tube temperature sensor error
Check cable connection
T02 Ai Error
Heated hose temperature sensor error
Check cable connection
T03 Ai Error
Heated valve temperature sensor error
Check cable connection
T10 Ai Error
Water tank temperature sensor error
Check cable connection
T11 Ai Error
Water temperature forward temperature sensor error
Check cable connection
T12 Ai Error
Water temperature return temperature sensor error
Check cable connection
F12 Ai Error
Flow sensor error
Check cable connection
-1S3 Main supply
General power OFF
Check Main switch -1S3 in control cabinet
-35Q3 Fuse M1 motor
Power supply for frequency drive
Check circuit breaker -35Q3 in control cabinet
-35Q19 fuse M1 Break
Power supply for M1 Motor break
Check circuit breaker -35Q19 in control cabinet
-10T8 24V CC
Fuse for control cabinet 24VDC
Check fuse -10T8 in control cabinet
-10T2 24V Emergency stop
Fuse for emergency power 24VDC
Check fuse -10T2 in control cabinet
-10T14 24V Retractor
Fuse for retractor power 24VDC
Check fuse -10T14 in control cabinet
-11T12 24V PC
Fuse for probe cooling power 24VDC
Check fuse -11T12 in control cabinet
-50S15 Service Mode
Switch for Service/Operation Mode
Shift to Operation mode
-45Q3 H1 fuse
Power supply for Probe heating
Check circuit breaker -45Q3 in control cabinet
-46Q3 H2 fuse
Power supply for Heated hose
Check circuit breaker -46Q3 in control cabinet
-47Q3 H3 fuse
Power supply for Heated valve
Check circuit breaker -47Q3 in control cabinet
-101S2 M1 safety switch
Switch for Probe Safety ON/OFF
Shift to ON
-152F3 M20 fuse
Power supply for Heat Exchanger Fan M20
Check circuit breaker -152F3 in Water cooling control cabinet
-153F3 M21 fuse
Power supply for Heat Exchanger Fan M21
Check circuit breaker -153F3 in Water cooling control cabinet
-154Q3 H10 fuse
Power supply for Water Tank Heat H10 (Option)
Check circuit breaker -154Q3 in Water cooling control cabinet