Navigator 600 Phosphate
Multi-stream
5 Configuration
38
IM/NAV6P/MS–EN Rev. F
5.8 Communications
There are three tabs in the I/O modules screen:
5.8.1 Ethernet
Configures the way that the analyzer can be accessed via an Ethernet network – see Appendix E, page 76.
Note.
Changes to the 'IP-address', 'Subnet mask' and 'Default Gateway' parameters are implemented only after the analyzer
has been restarted. Change the addressing parameters, exit and save the configuration, then power down and restart the
analyzer.
Fields
Description
IP-address
The IP-address to be assigned to the analyzer. The IP address is used by the TCP/IP protocol to
distinguish between different devices. The address is a 32 bit value expressed with four values (0 to
255), each separated by a period (.).
Subnet mask
The subnet mask is used to indicate which part of the IP address is for the network ID and which is
for the host ID. Set each section that is part of the network ID to 255.
For example, 255.255.255.0 indicates first 24 bits are for the network ID.
Default Gateway
The IP address for the 'Default gateway' (for example, router, switch) required to communicate with
other networks.
The default setting is '0.0.0.0'
FTP user 1 to FTP user 4
Enables up to four users to access the analyzer via the internet. A dialog box is displayed with four
options:
User name
– the name of user granted FTP access (up to 12 characters).
Password
– the password required for FTP login (up to 12 characters).
Access Level
– toggles between 'Full' or 'Read-only' access.
Remote Operation
– toggles between 'None', 'Operator' or 'Configuration'.
Note
. If a user is given full access via FTP, that user is able to select from the saved
configuration files in the analyzer.