Navigator 600 Phosphate
Multi-stream
5 Configuration
IM/NAV6P/MS–EN Rev. F
35
5.6 Current Outputs
There are six tabs in the Current Outputs screen, one for each output:
Fields
Description
Output Source
The 'Output Source' field has several options:
None. No other fields are visible
Stream 1 to Stream 6 – Additional fields appear
Output Range
The high and low limits for the output range. Both values can be set independently (0 to 15 ppm).
If the difference between the zero and span is too small, the output is very noisy.
Output Type
The electrical high and low limits (0 to 22 mA).
For example, if the output range is set to 0 to 15 ppm and the 'Output Type' to 4.00 to 20.00 mA, at
0 ppm the output is 4.00 mA and at 15 ppm the output is 20.00 mA.
Out of Sample Ind.
Out of sample indicator. If set to 'Yes', the output goes to the default output value when an out of
sample condition occurs for the selected stream source.
Default Output
The output value used when an out of sample condition occurs and 'Out of Sample Ind.' is set to
'Yes' (0 to 22 mA).