Navigator 600 Phosphate
Multi-stream
5 Configuration
IM/NAV6P/MS–EN Rev. F
33
5.5 Alarm Relays
Fig. 5.3 High / Low Process Alarms
Fig. 5.4 High / Low Latch Alarms
Fig. 5.5 High / Low Annunciate Alarms
Hysteresis
Concentration
Alarm On
Alarm switches
off automatically
Alarm Off
Trip Point
High Alarm Action
Low Alarm Action
Hysteresis
Concentration
Concentration
Alarm On
Alarm switches
off automatically
Concentration
Hysteresis
Alarm Off
Trip Point
Alarm On
Alarm Latched
Alarm
Off
Alarm
acknowledged
by operator
Earliest point that
the alarm can be
acknowledged
Earliest point that
the alarm can be
acknowledged
Hysteresis
Alarm On
Alarm Latched
Concentration
Alarm
acknowledged
by operator
High Latch Alarm Action
Low Latch Alarm Action
Concentration
Alarm
acknowledged
by operator
Concentration
Alarm switches
off automatically
Alarm Off
Trip Point
Alarm
acknowledged
by operator
Hysteresis
Alarm On
Alarm switches
off automatically
High Annunciate Alarm Action
Low Annunciate Alarm Action
Hysteresis
Alarm On