Navigator 600 Phosphate
Multi-stream
6 Start-up
IM/NAV6P/MS–EN Rev. F
41
6.2 Performing a Baseline Calibration
To perform a Baseline calibration:
1. Press the
key and select 'Calibration & Maintenance'
in the menu.
2. Select the 'Calibration' button and press the
key.
3. Change the 'Secondary Calibration' value to match the
concentration of the secondary calibration solution being
used.
4. Start a 'Baseline Calibration'.
The 'Baseline Calibration' window is displayed:
Note.
Before starting the initial Baseline calibration, it is
important to check the reagent lines are fully purged
and there is no air flowing through the reaction block.
Check there are no air bubbles in the cuvette by
checking the stability of the 'Measure Detector mV'
signal in the 'Measurement Status' / 'Diagnostic
Information' page.
If the signal is moving by more than 20 mV over a 30
second period, air may still be present in the cuvette.
Initiate a manual clean sequence.