background image

UAV- M I CROPORTA B LE  A N A LY ZE R S  |

 G L A 1 3 3 SER I E S | U M/G L A 1 3 3 - EN R E V. B

27

Figure 38  Serial Logger Menu

 . . .7  User Interface Operation

Serial Logger

Click the 

Serial Logger

 tab to access the 

Serial Logger menu .  

Here, you can change how data reported at the RS-232 ports 
are configured .  (Figure 38)

 NOTE

Use a null modem serial cable to connect the analyzer serial 
port to an external computer .

 NOTE

The 

OUTPUT

 tab corresponds to the 

Active Listeners 

ICOS 

tab .

Active Listeners (OPT-DATALOG Option)

Select which serial port(s) are available to stream data: 

ICOS

ANEMOMETER

, or

 DJI

 (or 

MAVLINK

) . Check the adjacent box to 

activate the port; uncheck to de-activate .

Serial Port Settings

Use the 

DJI

 (or 

MAVLINK

), 

ANEMOMETER

, and 

OUTPUT

 tabs to 

configure the selected serial port(s) that stream data . Figure 38 
shows the 

Serial Logger menu with the 

OUTPUT 

tab selected .

Standard settings are provided for:
•  Baud rate
•  Parity
•  Stop bits
•  Data bits
•  Timeout

The actual rate of the serial output is equal to the Logged File 
Rate (that is, 1 Hz) divided by the Rate specified in the 

Rate 

Control Adjustment panel .

You can also name the port .

Additional settings on the 

OUTPUT

 tab:

•  Active

 – check this box to activate the port for data output . If 

unchecked, no data are output from the port .

•  Append checksum

 – check this box to include serial port data 

columns in the output file .

Additional setting and display on the 

ANEMOMETER

 tab:

• 

Parser Settings – select the method to parse raw serial data 
into the correct format for data output .

• 

Raw data – displays the raw data output from the device on 
the selected serial port .

Parsed Debug

This panel displays the serial data stream .

•  Enable live debug

 – check this box to enable the data stream 

from the port . If unchecked, no data appear in the table .

•  Clean debug

 – click to clear the table of data .

•  Go to the table bottom

 – check this box to scroll to the 

bottom of the table .

 

Save or Restore Configuration

To save the configuration, click 

Save 

(at the bottom of the 

screen), or 

Restore

 to return the configuration to the default 

setting .

Содержание GLA133 Series

Страница 1: ... ABB MEASUREMENT ANALY TICS USER MANUAL GLA133 Series UAV Microportable analyzers ...

Страница 2: ...as Inlet 12 Power 12 Data Interface 12 Plumbing Diagram 12 7 User Interface Operation 13 Malware Protection 13 Set Up Communication 13 WiFi Setup 13 VNC Server First Time Activation 15 VNC Viewer User Interface 15 Log In 16 Security Button 17 Change User Password 17 Add New User 18 Log Out 18 Control Bar 18 Display 19 Numeric Display 19 Spectrum Display 19 Timechart Display 20 Alarm Status Display...

Страница 3: ...Daily Operation Checklist 33 Recommended Maintenance 33 Mirror Ringdown Time and Maintenance 34 Other Components 34 Fuse Replacement 34 Analyzer Cover Maintenance 35 Appendix A Spare Parts and Kits 36 Mirror Cleaning Kit 36 Spare Parts Kit Pump and Filters 36 Spare Filters 36 Spare Analyzer Shroud 36 Spare Router 36 Appendix B Packaging and Shipping the Analyzer 36 Contents ...

Страница 4: ...e intrusion leakage and or theft of data or information ABB and its affiliates are not liable for damages and or losses related to such security breaches any unauthorized access interference intrusion leakage and or theft of data or information Patent The analyzer technology is protected by patents 7 468 797 6 839 140 6 795 190 6 694 067 Copyright 2023 ABB Inc All rights reserved No information co...

Страница 5: ...ata sampling rates and transfer data via WiFi on the device that has established and secured a link with the analyzer Even though this user manual provides additional information on the hardware components and their particular functions it is recommended to have qualified ABB service personnel resolve any issues with the analyzer The following items are provided as standard scope of supply Table 1...

Страница 6: ...in a manner not specified by the manufacturer it may impair the protection provided by the equipment This makes the manufacturer s warranty null and void The device must be used by properly trained personnel when measurements in ambient air are required Any use with background gases presenting toxic corrosive and or explosive properties is not intended and may result in serious injuries Sound Ther...

Страница 7: ...oints According to the WEEE Directive 2002 96 EC only products used in private applications may be disposed of at municipal garbage facilities Proper disposal prevents negative effects on people and the environment and supports the reuse of valuable raw materials Labels The following three labels are on or in the analyzer to identify hazardous areas The first two labels are located on the carbon f...

Страница 8: ...nterlock switch to open the case cover during analyzer operations can cause serious bodily injury Even though the analyzer provides a second layer of protection such as a laser cover to prevent the user from the invisible laser beam or any secondary reflection from the laser on a reflective surface it is not recommended to modify the analyzer to operate in an unsafe condition Electrical Hazards Th...

Страница 9: ...ly stored at temperatures between 4 C and 50 C 40 F and 122 F Table 5 lists and describes the safety icons on ABB shipping boxes Follow these instructions when transporting and storing boxed analyzers Table 5 Transportation and Storage Icon Descriptions Icon Meaning Store your analyzer in a sheltered dry area Do not let the box get wet Transport and store the analyzer box with the arrows on the bo...

Страница 10: ...22 45 cm 8 84 in x Height 10 61 cm 4 18 in see Figure 2 Physical Weight 3 kg 6 6 lbs max Power Source External DC power supply via 120 240 V AC outlet or UAV batteries Power Input 11 30 V DC 35 W GLA131 GGA 27 W GLA131 GPC CH4 Operating Current 5 A max Fuse 5 A 32 VDC automotive fuse Fast blow fuse Altitude 2000 m 6562 ft Operational Temperature 5 to 45 C 41 to 113 F Humidity 99 non condensing Ins...

Страница 11: ... avoiding condensation on surfaces Maximum Impact Force Allowable 10 G Parameters Specifications Measurement Accuracy 1 of NIST Traceable Gas Standards Sampling Rate 0 01 10 Hz Environmental Requirements Table 8 defines the environment requirements of the analyzer during operational and non operational conditions Operation outside of these requirements may cause damage to the analyzer Table 8 Envi...

Страница 12: ...ness Power can be supplied using the 120 240 V power cable supplied with the analyzer or from the UAV batteries if applicable refer to the HoverGuard Solution System Manual document number 3KXG167006R4601 Data Interface USB ports used to transfer data to a USB memory device or connect a USB keyboard and mouse WiFi External Antenna port allows for wireless connection to the analyzer via an external...

Страница 13: ...for setup instructions WARNING To maximize cybersecurity users are advised to change all passwords from their default value WiFi Setup 1 Connect the external antenna provided with the analyzer to the WiFi external antenna port see Figure 4 2 Power on the analyzer by pressing the power button on the connector panel to the On position 3 On the tablet cell phone or laptop go to the WiFi symbol to acc...

Страница 14: ...least six characters long d Click Save 7 Change the Wireless Key WiFi password a On the left side of the screen click Network Wireless To the right of the SSID click Edit Click on the Wireless Security tab and click to show the Key Figure 9 Figure 8 Admin Password Panel Figure 9 Wireless Settings Panel Figure 10 Save New WiFi Password 7 User Interface Operation b Input the new WiFi password in the...

Страница 15: ...wireless remote control connection between the analyzer and the interface The analyzer uses VNC Viewer by RealVNC Limited for communication between the user and the analyzer VNC Viewer Figure 12 can be downloaded from Play Store for Android devices App Store Apple OS devices or directly from the RealVNC Internet site for PC or Mac installation www realvnc com The following procedure is for a Windo...

Страница 16: ...on shown in Figure 13 Check the Windows Start menu application list b Click on the VNC icon to execute the application the screen shown in Figure 14 appears c Enter the IP address of the analyzer Figure 7 The format of the IP address is 192 168 1 XXX where XXX consists of three digits that vary from analyzer to analyzer d When prompted enter the password lgrvnc The analyzer s software display shou...

Страница 17: ... SSH password also changes WARNING If you forget the new password it is not recoverable Security Button Once logged in the Security button also lets you do the following Change a user password Add a new user Log out of the system Change User Password Upon commissioning of the analyzer all passwords should be changed Table 10 lists the default passwords Table 10 Default Passwords Default Passwords ...

Страница 18: ...add a new user perform the following 1 Click the Security button 2 Click Add User The Add User dialog box shown in Figure 19 appears 3 Enter a username in the User Name field 4 Enter a password in the Password field 5 Re enter the new password in the Confirm Password field 6 Select a group from the Group Name drop down menu Admin Service or User 7 Click Add New User If you try to add a user and ar...

Страница 19: ...ing on the analyzer type In addition the analyzer color coded operational status is indicated for details on color coding refer to Alarm Status Display Figure 26 Spectrum Display The sample Spectrum display provides additional information on the measured gas the analyzer sensitivity level shown in the Transmitted Intensity diagram the absorption level a theoretical fit to the measured target gas e...

Страница 20: ...button in the Timechart display Figure 25 refreshes the screen to display the sampling data with a new starting point Alarm Status Display The Alarm Status display Figure 26 indicates the real time analyzer operational status The status uses traffic light color indication Green indicates no problem Yellow indicates the analyzer is out of spec or maintenance is soon required Red indicates there is ...

Страница 21: ...ors soon Warning Linelock Laser peak position is outside of control range contact customer support Warning Signal Power Laser power has degraded contact customer support Warning Maintenance Maintenance is needed on system soon No issue No warning alarm Rate Button The Rate button on the Control Bar changes the rate at which data are written to the log file Figure 27 displays the Rate Control pop u...

Страница 22: ...alyzer serial number Next set of characters represent the date yyyy mm dd Last set of characters give the data file name fxxxx txt where xxxx is the serial number The serial number counts upward to provide up to 10 000 unique file names each day If the analyzer is left in continuous operation a new data file is automatically created every 24 hours to keep data file sizes manageable Standard Data F...

Страница 23: ...ates To access the data files for a specific date double click the folder Each file from that day is displayed in chronological order Figure 30 Each file is a single zipped txt file using the following convention DJI_Analyzer Serial Number_YYYY MM DD_f0000 txt zip Examples of files in the daily directory are shown in Figure 30 Archive Directory The local hard drive Figure 29 creates an archived fo...

Страница 24: ...RS 232 ports are configured Refer to the Serial Logger section in page 27 Security Change security settings including changing passwords adding deleting and assigning users and returning the system to default settings Refer to the Security section Service Technicians can check on the analyzer status Time Files Click the Time Files tab to access the Time Files menu Figure 34 This screen lets you ad...

Страница 25: ...yzer without shipping it back to the factory Before performing a calibration on the analyzer have the following information available Traceable regulated gas type Traceable regulated gas type concentration WARNING It is not recommended for inexperienced users to perform a calibration without first contacting Technical Support To perform a gas calibration connect the traceable regulated bottle gas ...

Страница 26: ... the theoretical target To achieve this adjust the voltage driving the laser temperature to move the laser to the operating wavelength To manually adjust the laser wavelength 1 Click the Setup button on the Control Bar Figure 21 2 Select the Laser Adjust tab at the top of the screen Figure 36 3 Select the Disable Laser Frequency Lock checkbox to allow manual control of the laser 4 Adjust the Laser...

Страница 27: ...bits Timeout The actual rate of the serial output is equal to the Logged File Rate that is 1 Hz divided by the Rate specified in the Rate Control Adjustment panel You can also name the port Additional settings on the OUTPUT tab Active check this box to activate the port for data output If unchecked no data are output from the port Append checksum check this box to include serial port data columns ...

Страница 28: ...g the connection to the VNC Server will not be encrypted as shown in Figure 41 7 User Interface Operation Security Click Setup on the Control Bar Figure 21 then click the Security tab to access the Security screen Figure 39 This screen lets you change security settings including changing passwords adding deleting and assigning users and returning the system to default settings Returning to default...

Страница 29: ...4 To proceed click Yes The VNC server is enabled and you can connect to the analyzer via the VNC interface You can also disable the connection Click Change VNC Password Figure 42 5 The Change VNC Password dialog box appears Figure 43 6 Enter the current VNC password in the Current Password field The default VNC password is lgrvnc 7 Enter the new password in the New Password field 8 Re enter the ne...

Страница 30: ... Change Samba Password Dialog Box Manage User Accounts User Levels The analyzer is shipped with one Admin user lgr which cannot be deleted Refer to Table 13 for user level access Table 13 User Level Access Admin Service User Display Screens Numeric Spectrum Time Chart Alarm Status Main Buttons Rate File Transfer Setup Exit Setup Panels Time Files Calibration Laser Adjust MIU Analog Output Serial L...

Страница 31: ...w User If you try to add a user and are in the User group instead of the Admin or Service group a dialog box appears and you must cancel the action 7 User Interface Operation Figure 46 Group User Dialog Box NOTE Only users belonging to the Admin and Service groups can add new users NOTE Only users belonging to the Admin and Service groups can delete users Delete User To delete a user perform the f...

Страница 32: ...missioned or released from active service This feature requires the lgr user Linux account and does the following Sets default passwords in the user management group Removes all measurement data and created users Disables the VNC server To set the OA ICOS default settings perform the following 1 From the Setup screen click the Security tab 2 Click Clear Icos Data The Clear Icos Data dialog box sho...

Страница 33: ...lyzer must be performed only by certified service personnel trained on servicing this instrument User adjustments inside the analyzer are not necessary or recommended by the manufacturer Frequency Task Every 1 2 days On the Spectrum display verify that the spectrum is correct The spectrum should appear as shown in Figure 23 Become familiar with the normal appearance of the spectrum the best way of...

Страница 34: ...ed Other Components The inline particulate filter requires replacement once a year or when the measured data profile becomes thinner in the Spectrum display screen GOF gets worse refer to the Spectrum Display section for details and or when the gas pressure reading on the Control Bar increases Figure 50 Solid state drive SSD maintenance It is recommended to check the amount of memory space availab...

Страница 35: ...ve the analyzer back to the original location 12 Re connect the analyzer power plug Figure 53 Fuse with Fuse Cover Removed NOTE Use Threadlocker Loctite 243 for the screws used to attach the shroud to the baseplate Analyzer Cover Maintenance The shroud is designed to minimize the weight of the analyzer cover while also protecting the analyzer components and eliminating external light sources from ...

Страница 36: ...abb com 2 Disconnect all connections from the analyzer Return the external antenna with the analyzer 3 Cover the gas inlet 4 Carefully place the analyzer in the plastic bag then tape the bag shut to protect the analyzer from dust and other particulate matter during shipping Place the external antenna in a small plastic bag and tape to the analyzer plastic bag 5 The foam packaging material may be n...

Страница 37: ...k of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB ABB 2023 3KXG167004R4601 ABB Inc Industrial automation 3400 Rue Pierre Ardouin Québec Québec G1P 0B2 Canada Tel 1 800 ...

Отзывы: