Chapter 10: Status Messages, Troubleshooting
42/24-10 EN Rev. 9
Caldos25, Caldos27, Magnos206, Magnos27 Problems
Temperature Problem
Faulty temperature
sensor or heater
connections
•
Check the connecting lines and plugs.
•
Check the line seating in the insulated jackets.
Defective thermal link
•
Check thermal link continuity and replace if necessary (see page 9-3 for
instructions).
Leaking thermostat
or open purge gas
connections
•
Check the seal integrity between the thermostat chamber and the mounting
flange; tighten mounting bolts or replace O rings as needed.
•
Seal the analyzer purge gas inlet and outlet with sealing connectors.
•
Check the purge gas flow (maximum operating level of 20 l/h; positive pressure
p
e
= 2 to 4 hPa) and reduce as needed.
Unstable Readings
Gas path leakage
•
Check the integrity of the analyzer module gas paths and (if applicable) of the
lines to the gas module (see page 9-2 for instructions).