10-10
Chapter 10: Status Messages, Troubleshooting
42/24-10 EN Rev. 9
Status Messages,
continued
No. Text
O
E M F Reaction / Remark
308 A computer error occurred
during calculation of the
measurement value.
x x Notify service.
309 The temperature regulator is
defective.
x See the status message from the applicable temperature
detector
310 Temperature correction turned
off for this component
because of invalid
temperature measurement
value.
x See the status message from the applicable temperature
detector
311 The pressure regulator is
defective.
x x See the status message from the applicable pressure
detector
312 Pressure correction turned off
for this component because of
invalid pressure measurement
value.
x See the status message from the applicable pressure
detector
313 Cross-sensitivity correction is
impossible for this component
because the correction value
is invalid.
x See the status message from the applicable correction
detector
314 Carrier gas correction is
impossible for this component
because the correction value
is invalid.
x See the status message from the applicable correction
detector
Auxiliary Detector
315 No new measurement values
from analog/digital converter.
x Notify
service.
316 Measurement value exceeds
the analog/digital converter
value range.
x Notify
service.
317 A computer error occurred
during calculation of the
measurement value.
x Notify
service.
Uras
318 No new measurement values
from analog/digital converter.
x x Notify service.
Caldos, Magnos
319 The measurement bridge is
improperly balanced.
x x Notify service.
320 The measurement amplifier
offset is too high.
x x Notify service.
Continued on next page