A-2-8
Appendix 2: Analyzer Module Operating Specifications
42/24-10 EN Rev. 9
Electrical Safety
Test
per EN 61010-1: 2001
Protection Class
Central unit with electronics module (power supply):
I
Analyzer module without electronics module (power supply):
III
Overload Category /
Pollution Level
Electronics module power supply:
III / 2
Analyzer module power supply:
II / 2
Signal inputs and outputs:
II / 2
Safe Isolation
The electronics module power supply is galvanically isolated from other circuits by
means of reinforced or double insulation. Operational low voltage (PELV) on low-
voltage side
Electromagnetic Compatibility
Noise Immunity
Tested to EN 61326-1: 2006.
Inspection severity: Industrial area, fulfills at least the rating “continuously
monitored operation” to Table 2 of EN 61326.
Emitted Interference
Tested to EN 61326-1: 2006, EN 61000-3-2 : 2006 and EN 61000-3-3 : 1995 +
A1: 2001 + A2 : 2005.
Limit value class B for interference field strength and interference voltage is met.