7-C-8
Chapter 7: Gas Analyzer Configuration
42/24-10 EN Rev. 9
Replacing a System Module
Removing and
Reinstalling an
Existing System
Module
As a rule, when an existing system module is removed and reinstalled (e.g. after a
repair) it does not require any setup.
When the system module is reconnected to the system bus it is automatically
recognized and its configuration is automatically stored. For automatic recognition
to take place the gas analyzer must be in measurement mode.
CAUTION!
When an existing system module is replaced with another system module the
function “DELETE” should not be used to delete the old system module. This
function would irretrievably delete the parameter settings and function block
configuration of the old system module!
In order to preserve the parameter settings and function block configuration
of the old system module when replacing a system module, the function
“CHANGE” must be used.
•
The type and configuration of the new system module must match the charac-
teristics of the old system module.
•
When an existing I/O module has been replaced with an I/O module of the same
type, the new I/O module is automatically recognized by the gas analyzer and its
configuration is not necessary.
Step Action
1 Select
the
Set Up System Modules
menu item.
The list of the system modules existing in the system is displayed.
2
Select the system module (analyzer module or I/O device) to be re-
placed and reconfigured. In the list this system module will be shown
with an
Unknown
or
Error
status.
Press the
CHANGE
softkey.
3
Do not press the
DELETE
softkey! This will irretrievably
remove the system module’s parameter settings and function
block configuration!
4
Enter the 14-digit serial number for the new system module.
5
In the list the new system module will now be shown with a
Replaced
status.
Replacing an Existing
System Module
(Analyzer Module or
I/O Device) with
Another System
Module
6
Save the configuration change by pressing
ENTER
or discard it by
pressing
Back.