16
3.
CHANGING TYPE
3-1 Changing from SV-A (Stand type) to SV-H (Portable type)
In the case that you have purchased an SV-A series (stand type), and you need to use it as a
portable type, make the following changes.
Note: When removing or tightening the screws,
use the screwdriver that is provided
with the viscometer.
1 Remove the four screws from the
rear side of the sensor unit’s slider.
Using the four screws removed,
attach the sensor unit to the handle.
Change the function setting of
“
5vtyp
”
from
“
atype
”
to
“
Htype
”
as follows:
* Even if the function setting of
“
5vtyp
”
is not changed, you can still use the viscometer. However,
the portable functions,
H-fnc
,
5t-b
,
Hld-t
,
Cp
,
bep
,
Cp Hi
and
Cp lo
will not be available.
2
Insert the AC adapter plug into the AC adapter
jack located on the rear side of the display unit.
Insert the other end of the AC adapter plug into an
electrical outlet.
3 Press the ON:OFF key to turn the power on.
4 While in the standby mode, press and hold the MODE
key for two or more seconds. The viscometer will
enter the function mode
5vtyp
.
5
Press the PRINT key to select this item.
AC adapter cable
Display unit
rear side
(Standby mode)
Press and
hold
Continued on next page
Screws
Screws
Handle
Screwdriver provided
with the viscometer