9-4 Data Output Format Examples (For all models)
9-4-1 A&D Standard Format
Used with the printer MODE 1 or MODE 2 when the optional compact printer AD-8121B is
connected. Only the viscosity value is output.
SV-1A/SV-1H output format example
Viscosity
unit
Display Output
format
Remarks
L
LLLL
mPa
・
s
OL,-99999999mPs
Below measuring range error
0.30mPa
・
s
ST,+00000.30mPs
10.0 mPa
・
s
ST,+00010.00mPs
The digit of 0.01mPa
⋅
s is always zero.
100 mPa
・
s
ST,+00100.00mPs
The digits of 0.01mPa
⋅
s and
0.1mPa
⋅
s are always zero.
1.00 Pa
・
s
ST,+01000.00mPs
For 1000mPa
⋅
s or greater, the
displayed unit is Pa
⋅
s, but the
output unit remains mPa
⋅
s.
The digits of 0.01mPa
⋅
s, 0.1mPa
⋅
s
and 1mPa
⋅
s are always zero.
m
mPa
・
s
H Pa
・
s
OL,+99999999mPs
Above measuring range error
L
Pa
・
s
OL,-99999999Pas
Below measuring range error
0.0003 Pa
・
s
ST,+000.0003Pas
0.0100 Pa
・
s
ST,+000.0100Pas
0.100 Pa
・
s
ST,+000.1000Pas
The digit of 0.0001Pa
⋅
s is always
zero.
1.00 Pa
・
s
ST,+001.0000Pas
The digits of 0.0001Pa
⋅
s and
0.001Pa
⋅
s are always zero.
Pa
・
s
H
Pa
・
s
OL,+99999999Pas
Above measuring range error
L cP
OL,-99999999 cP
Below measuring range error
0.30 cP
ST,+00000.30 cP
10.0 cP
ST,+00010.00 cP
The digit of 0.01cP is always zero.
100 cP
ST,+00100.00 cP
The digits of 0.01cP and 0.1cP are
always zero.
10.0 P
ST,+01000.00 cP
For 1000 cP or greater, the displayed
unit is P, but the output unit remains cP.
The digits of 0.01cP, 0.1cP and
1cP are always zero.
C
cP
H P
OL,+99999999 cP
Above measuring range error
L P
OL,-99999999
P
Below measuring range error
0.0030 P
ST,+000.0030
P
0.100 P
ST,+000.1000
P
The digit of 0.0001P is always
zero.
1.00 P
ST,+001.0000
P
The digits of 0.0001P and 0.001P
are always zero.
10.0 P
ST,+010.0000
P
The digits of 0.0001P, 0.001P and
0.01P are always zero.
P
H P
OL,+99999999
P
Above measuring range error
: Space (ASC 20h)