![A&D SV-1A Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/aandd/sv-1a/sv-1a_instruction-manual_3103626020.webp)
18
2 Remove the attachment screw from the
base. Install the X-Y-Z stage along the guide
ribs from forward of the base. Secure the
X-Y-Z stage using the screw removed.
(Confirm the direction of the X-Y-Z stage’s
knobs.)
3 Remove the four screws from the
rear side of the sensor unit’s
handle.
Using the four screws removed,
attach the sensor unit to the slider.
4 Raise the lever an d install the sensor unit along
the guide channels of the supporting post’s
sides.
At an appropriate height, secure the sensor
unit to the supporting post by lowering the
lever.
Change the function setting of
“
5vtyp
”
from
“
Htype
”
to
“
atype
”
as follows:
* Even if the setting of
“
5vtyp
”
is not changed, you can still use the viscometer. However, the
portable functions,
H-fnc
,
5t-b
,
Hld-t
,
Cp
,
bep
,
Cp Hi
and
Cp lo
remain available.
Slider
Screws
Screwdriver
Screw
Install from forward
X-Y-Z stage
Guide channel
Serial number
Lever
Handle the sensor
plate with care.