47
9 When the measurement of the first point has
completed, clean the sensor plates,
temperature sensor and protector and
prepare the second standard viscosity fluid.
10 Place the second standard viscosity fluid in
the sample cup. Press the START key to
start the measurement of the second point.
11 After the measurement, wait for the display
to become stable and press the PRINT key.
The measurement value blinks and is ready
to be corrected.
12 Correct the value using the following keys:
MODE key
Switches the blinking digits.
START key
Changes the setting of a
blinking digit.
STOP key
Moves the decimal point.
13 Press the PRINT key to confirm the
correction value.
14 To exit the calibration mode:
With "
end
" being displayed, press the PRINT
key again. The calibration data is saved and the
display returns to the standby mode.
To correct the value:
With "
end
" being displayed, press the
STOP key and correct the value.
15 Measure the viscosity of the two standard
viscosity fluids used. Confirm that the
viscometer displays the similar value as the
entered correction value for each fluid.
From the previous page
Set standard viscosity fluid
(Second point measurement mode)
When the data
is stabilized
(Correct the second point value)
(End of two-point calibration)
To correct
(Standby mode)
(Confirm the value
of the first fluid)
(Confirm the value
of the second fluid)