35
5 After measurement, the buzzer sounds and the displayed
value blinks, then holds.
Note
When the values do not stabilize because of hand
movement, set the condition function that levels the
large variation in digits (refer to page 54) and the
function that do not display the large variation in digits
(refer to page 55) if necessary.
With the portable type, you can use the auto hold
function and timer mode function (refer to page 57).
6 To stop the measurement, press the STOP key.
To go on to the next measurement, set the sample and press
the START key.
When the viscosity value is below the measuring range,
Sl TOP
is displayed.
When the viscosity value exceeds the measuring range,
SH TOP
is displayed.
If the sensor plates are not in vibrating motion for about 20 seconds, for example, when
the sample coagulates, the measurement will stop automatically.
∗
1 When using the graphing program RsVisco, of the AX-SV-53-EX software set, click the
START button of the RsVisco program to start a measurement.
∗
2 While the measurement is being performed using the graphing program RsVisco of the
AX-SV-53-EX software set or data are being output continuously (function setting "
prt 2
"
or SIR command), the data hold mode using the HOLD key is not available.
Below is an example of the SV-10H at shipment.
(Standby mode)
(Measurement mode)
(Data hold
after measurement)