95
Display
Error code
Description
mode.
EC,E21
Calibration weight error
The calibration weight is too light. Confirm the calibration
mass value. Press the CAL key to return to the weighing
mode.
Overload error
A sample beyond the balance weighing capacity has been
placed on the pan. Remove the sample from the pan.
Weighing pan Error
The mass value is too light.
Confirm that the weighing pan is properly installed and
calibrate the balance.
Sample mass error
The balance can not store the sample for the counting mode
or for the percent mode because it is too light.
Use a larger sample.
Unit mass error
The sample unit mass for the counting mode is too light.
Storing and using it for counting will cause a counting error.
Add samples to reach the specified number and press
the PRINT key. Pressing the PRINT key without
adding samples will shift the balance to the counting
mode. But, for accurate counting, be sure to add
samples.
Clock battery error
The clock backup battery has been depleted.
Press any key and set the time and date. The clock and
calendar function works normally as long as the AC adapter
is connected to the balance. If this error appears frequently,
contact the local A&D dealer.
Power supply voltage fault
The voltage supplied from the AC
adapter is abnormal.
Please check if the problem is the AC adapter (TB248)
supplied with the balance.
Full memory
The maximum number or stored weighing values has
been reached. In order to store more weighing values, it
is necessary to delete the data. Refer to "11.Data
Memory"
Full memory
The stored calibration history has reached 50 instances.
If more is stored, the old history will be deleted. Refer to
"11.Data Memory"
EC,E00
Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC,E01
Undefined command error
(
)
(
)
Alternate
Blink
Blink
Alternate