76
In the case of using 0.1g model
Flow rate
(mL/min)
Response priority ←
Ct setting →
Accuracy priority
1sec 2sec 5sec 10sec 20sec 30sec 1min 2min 5min
10min
0.01
0.02
0.05
0.1
0.2
0.5
1
2
5
10
20
50
100
Example) When measuring the flow rate of 0.02 ml/min
Set the calculation time in 5minutes or more
In the case of using 0.01g model
Flow rate
(mL/min)
Response priority ←
Ct setting →
Accuracy priority
1sec
2sec
5sec
10sec 20sec 30sec 1min 2min 5min
10min
0.1
0.2
0.5
1
2
5
10
20
50
100
200
500
1000
Example) When measuring the flow rate of 20 ml/min
Set the calculation time in 5 seconds to 30 seconds
In the case of using 0.1g model
Flow rate
(mL/min)
Response priority ←
Ct setting →
Accuracy priority
1sec
2sec
5sec
10sec 20sec 30sec 1min 2min
5min
10min
1
2
5
10
20
50
100
200
500
1000
2000
5000
10000
Example) When measuring the flow rate of 2000 ml/min
Set the calculation time in 1 second to 5 seconds