22
Processing mark
Counting Mode Using The ACAI Function
The ACAI is a function that improves the accuracy of the unit mass automatically by increasing the number of
samples as the counting process.
ACAI: Automatic Counting Accuracy Improvement
After registering unit mass of "Step 6", proceed to the following "Step 8".
8. If a few more samples are added, the processing indicator turns on. To
prevent an error, add three or more. The processing indicator does not turn
on if overloaded. Try to add the same number of samples as displayed.
9. The balance re-calculates the unit mass while the processing indicator is blinking. Do not touch the
balance or samples on the pan until the processing indicator turns off.
10. Counting accuracy is improved when the processing indicator turns off.
11. Each time the above operation is performed, a more accurate unit mass will be obtained. There is
no definite upper limit of ACAI range for the number of samples exceeding 100. Try to add the
same number of samples as displayed.
12. Remove all the samples used in ACAI and proceed with the counting operation using the improved
unit mass.
Note ACAI will not function on the unit mass entered using the keys, or
digital input mode.
Storing the unit mass
By using the data memory function, 50 instances of storing a sample
unit mass can be stored.
1. Set the function setting item "Data memory function ( ) " to
"Stores unit mass in counting ( ) " .Refer to "10.Function
Table".
2. The displayed "P
**
" is the selected unit mass registration number.
3. Press and hold the PRINT key to switch to the mode to change the
unit mass registration number.
RE-ZEROkey
Changes the registration )
MODEkey
Changes the registration number (-)
PRINTkey
Decides on the displayed registration number.
CALkey
Cancel the displayed registration number.
4. Multiple unit masses can be stored by registering them with different unit mass registration numbers.
Note
※
Unit weight can be read by ″
UN
:
mm
″command.
(
mm corresponds to P01 to P50 with 01 to 50.
)
※
The read unit mass can output by
″?
UW
″
command and can be changed by
″?
UW ″
command.
Note
※
ACAI cannot be used for the read unit mass.
Press and Hold